Sash Operation
The counterbalanced, anti-racking sash mechanism requires only a few pounds of force
to move the sash up or down. You can open or close the sash smoothly with one or two
hands positioned on either handle.
The sash position alarm and safety interlock system senses the sash position and acts
appropriately. An audible sash sound will occur when the interlock system senses the
sash is at operating height. The cabinet has been programmed to operate at a 9-inch
(229 mm) sash opening. When the Blower is ON, moving the sash above or below its
operating height will activate the audible and visual alarms. The audible alarm can be
temporarily muted for approximately five minutes by pressing the
[OK/MUTE]
button on
the Keypad. Moving the sash back to its operating position will reset the alarm. The
safety interlock system senses when the sash is closed and allows the optional UV
lamp to operate only when the sash is closed, to protect the operator from irradiation.
Blower Operation
The Cell Logic Type A2 biosafety cabinet can be configured to automatically turn on the
internal blower via Smart-Start. If this feature is enabled, raising the sash from closed to
operating height will automatically turn on the blower. If this feature is not enabled, raise
the sash to operating height, and press [BLOWER]. Either method will start the blower,
and the
Airflow Status
will read
Wait
for approximately 1 minute.
Light Operation
The Cell Logic Type A2 biosafety cabinet provides dual LED lamps for illuminating the
work surface. The cabinet can be configured to automatically turn on the LED lighting
via Smart-Start. If this feature is enabled, raising the sash from closed to operating
height will automatically turn on the LED lighting. If this feature is not enabled, raise the
sash to operating height, and press [LIGHT]. Either method will illuminate the LED
lighting.
Outlet Operation
The Cell Logic Type A2 biosafety cabinet provides two GFIC outlets with a maximum
total amperage draw of 5A (100-115v models) or 3A (230v models). Power to the
outlets can be activated or deactivated from the Keypad. To activate or deactivate
power to the outlets, press [OUTLET].
69