Download Print this page

Labconco Cell Logic A2 User Manual page 37

Biosafety cabinets
Hide thumbs Also See for Cell Logic A2:

Advertisement

Certification
Prior to use, a qualified certifier should certify a biosafety cabinet. Under normal
operating conditions, the cabinet should be recertified at least annually and when
relocated or serviced. The certifier should perform the following tests, as recommended
in NSF/ANSI Standard Number 49 in effect when the cabinet was manufactured:
• Downflow Velocity Profile Test
• Inflow Velocity Test
• Airflow Smoke Patterns
• HEPA Filter Leak Test
• Optional Canopy Alarm Test and Operation
• Vibration Test *
• Noise Level Test *
• Lighting Intensity Test *
*These tests are user comfort related tests and may be
omitted at the user's or certifier's discretion.
If you have any questions regarding certification agencies or help locating one, contact
Labconco's Product Service Department at (800) 821-5525 or +1 816-333-8811.
Detailed information for product certification is located in the Logic Technical Manual.
37

Advertisement

loading
Need help?

Need help?

Do you have a question about the Cell Logic A2 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Logic a2