System Reset Switch

The system reset switch removes power from the control board and microprocessor. It
is located on the top face of the electrical box on the rear, right side of the product. See
Figure 7-2. This switch will NOT remove all electrical power from the product. For
service operations, always disconnect the main electrical connection prior to removing
service panels.