Timer Operation (Optional)
There is a timer option that can be enabled and
disabled by a setting in the
Screen. The timer counts down from the
set point selected and chimes once the
time is zero. To adjust the status of the
timer, see
Settings>Configuration
section and Figure 6-16. To change the
timer chime, see
Settings>Display
section for more information.
If enabled, the timer can be found in the
bottom left corner of the
in Figure 6-1. To change the timer length, touch the
the timer, touch [Play/Pause]. To restart the time to the max set value, touch [Reset].
Reference Figure 6-16.
The timer can also be initiated, stopped or reset using the
The timer value must be set on the touchscreen display, but each press of the
key starts, stops and resets the timer (respectively).
Airflow Alert
While the blower is on, any sudden disruption to the airflow within the cabinet will trigger
an Airflow Alert. When an Airflow Alert activates, the display will automatically change
(Figure 6-17), and the audible alert tone will sound. The most common causes of an
Airflow Alert are:
• Blockage of the inlet grilles or exhaust outlet.
• Removal of the work surface or grille during operation.
Configuration
of this
of this
Home Screen
as shown
61
Play/Pause
Timer Value
[Timer
Value]. To start or pause
[TIMER]
Figure 6-16
Reset
key on the keypad.
[TIMER]
Need help?