Summary of Contents for Labconco Purifier Axiom 30441
Page 1
Original instructions Technical Manual and Specifications Purifier Axiom Type C1 Biosafety Cabinet Models 30441 30448 30461 30468...
Page 2
While the advice and information offered in this document are believed to be true and accurate at the date of going to press, neither the author, nor Labconco Corporation can accept any legal responsibly for any errors or omissions that may be made.
Page 3
Table of Contents Introduction Contacting Labconco Section One – Background Information QuickStarts What’s the password for the Axiom? How do I calibrate blower speed, filter life & airflow sensor? 10-11 How do I see the current settings? How do I change the configuration?
Page 4
Preparing the Biosafety Cabinet for operation Moving and Lifting the Cabinet Installing the Biosafety Cabinet on an Existing Work Surface Installing the Cabinet on a Labconco Base Stand Telescoping Base Stands Manual or Electric Lift Base Stands SoLo Electric Hydraulic Lift Base Stands...
Page 5
Section Five – Certification Procedures for the Axiom Accessing the Certifier’s Menu Adjusting the Inflow, Downflow and Filter Life Gauges 73-74 Axiom Filter Life Gauge Operation and Resetting for Local Conditions HEPA Filter Leak Test Information Supply HEPA Filter Leak Test If the Axiom is contaminated Supply HEPA Filter Leak Test Specifications If the Axiom is uncontaminated...
Page 6
Lamp Socket Replacement Electronics Module Access Electronic Module Component Identification Circuit Breaker Resetting a circuit breaker Circuit breaker specifications 117-18 Testing a circuit breaker Replacing a circuit breaker Power Supply Board Service Operations Fuse Testing & Replacement Display Board Service Operations Touchpad Connections Diagnostic Operations Motor Service Operations...
Page 7
As always, we at Labconco want to assist you in having a better understanding of our products and their operation; if you have any questions, or need additional information, please contact us.
Page 8
SECTION ONE – Background Information Quickstart The Axiom is a significant step forward in Biosafety Cabinet technology. While many internal components are the same or similar to our Logic+-platform products, its Quickstarts operation and certification are different. The that follow will help you show bold blue you how to answer the questions asked in each case.
Page 9
Quickstart What’s the password for the Axiom? Like the Logic, the Axiom models password protects the user from entering the Calibration menu to prevent them from adjusting the blower speed, filter gauge or the Service airflow sensor readings. New to the Axiom is a menu selection entitled Configuration ;...
Page 10
Quickstart How do I recalibrate the ✓ blower speed ✓ filter life gauge ✓ airflow sensor reading of the Axiom? 1. Press the Menu button. Services 2. Press the ▼ button until the option is highlighted (it will turn yellow). OK/Mute 3.
Page 11
Quickstart How do I recalibrate the ✓ blower speed ✓ filter life gauge ✓ airflow sensor reading of the Axiom? - Continued 10. If this is an initial certification or a recertification after supply Reset Filter Gauge to 100% HEPA filter replacement, select .
Page 12
Quickstart How do I see the Current Settings of the Axiom? Menu 1. Press the button. ▼ Services 2. Press the button until the option is highlighted (it will OK/Mute turn yellow). Press to enter the Services menu screen. Current Settings 3.
Page 13
Quickstart How do I change the Configuration of the Axiom? Menu 1. Press the button. ▼ Services 2. Press the button until the option is highlighted (it will turn yellow). OK/Mute 3. Press to enter the Services menu screen. Configuration 4.
Page 14
Quickstart How do I change the Configuration of the Axiom? – continued 9. The first screen allows you to decide if the cabinet is connected to an exhaust system or not. When the unit is configured for no exhaust connection, the Axiom will ignore the inlet relief valve sensor’s status.
Page 15
Quickstart How do I run a diagnostic on the sash position switches? Menu 1. Press the button. ▼ Services 2. Press the button until the option is highlighted (it will turn yellow). OK/Mute 3. Press to enter the Services menu screen. Diagnostic 4.
Page 16
Quickstart How do I run a diagnostic on the keypad, relays and contacts? Menu 1. Press the button. ▼ Services 2. Press the button until the option is highlighted (it will turn yellow). OK/Mute 3. Press to enter the Services menu screen. Diagnostic 4.
Page 17
Quickstart How do I run a diagnostic on the Inlet Relief Valve? Menu 1. Press the button. ▼ Services 2. Press the button until the option is highlighted (it will turn yellow). OK/Mute 3. Press to enter the Services menu screen. Diagnostic 4.
Page 18
How do I run a diagnostic on the Airflow Sensor? Menu 1. Press the button. 2. Press the ▼ button until the Services option is highlighted (it will turn yellow). OK/Mute 3. Press to enter the Services menu screen. Diagnostic 4.
Page 19
MyLogic 2. Press the ▼ button until the option is highlighted it will turn blue). OK/Mute 3. Press to enter the MyLogic menu screen. configure my Logic for use 4. Press the ▼ button until the option option is highlighted (it will turn white). 5.
Page 20
anything that changes the motor’s current speed, like: ✓ Blocking the grille. ✓ Raising the air temperature. ✓ Lowering the barometric pressure. will increase the motor’s speed, causing the Filter Life Gauge to decrease. Electronic dampening of the gauge will limit the change in readings to approximately 1%/minute. Axiom Biosafety Cabinets being installed at altitudes or operating at temperatures significantly different from those at the time of manufacture (900 +/- 500 ft above sea must...
Page 21
Motor HP Exhaust Motor/Blower Data Motor HP Fluorescent Lamp Data Fluorescent Lamps (2 each) F32T8 TL741 F32T8 TL741 F40T8 SP41 F40T8 SP41 Fluor. Lamp Labconco Part # 9721900 9721900 9721903 9721903 Fluor. Ballast Labconco Part # 3838100 3838100 3838100 3838100...
Page 22
Motor HP Exhaust Motor/Blower Data Motor HP Fluorescent Lamp Data Fluorescent Lamps (2 each) F32T8 TL741 F32T8 TL741 F40T8 SP41 F40T8 SP41 Fluor. Lamp Labconco Part # 9721900 9721900 9721903 9721903 Fluor. Ballast Labconco Part # 3838100 3838100 3838100 3838100...
Page 23
Measured at the exhaust transition sampling point, relative to atmosphere. IRV = Inlet Relief Valve. UV lamp Specifications For all models, the UV lamp number is G30T8. The Labconco part number is 1271300. ✓ For all 115/230 VAC models, the ballast assembly is Labconco part number 3829901. The ✓...
Page 24
IDENTIFYING AXIOM MODELS The Axiom platform Biosafety Cabinet can be identified readily by several external features, as shown in Figure 1-1: • Model number Stars with 304xxxxxx • Serial number beginning with 1503xxxxx. • Exhaust System Cover. • Segmented Work Surfaces. Figure 1-1 Exhaust System Cover Axiom Logo...
Page 25
Final Inspection and Test Report Form Obviously, you can get confirmation of the Biosafety Cabinet's model by contacting Labconco's Product Service Department, and providing us with the serial number. From this number, we can tell the date of production and its model.
Page 26
Axiom units’ model numbers define the configuration of the cabinet. The first five digits define the unit's configuration, while the last four are assigned based on the unit's options and electrical configuration. Thus, a model number of: 30448xxxx Labconco Product Group 3=Laminar Flow Products Indicates the Product Family 0= Logic...
Page 27
Axiom Data Plate The Axiom Data Plate contains much of the information required to certify the unit. Information on the plate includes nominal inflow and downflow velocity settings, a description of how to establish the downflow grid and the secondary inflow methodology are also included on the data plate, as shown below.
Page 28
Final Inspection and Test Report Form A copy of the Final Inspection Test Report is attached to the blower plenum cover, beneath the front dress panel. The report contains initial setup information on inflow and downflow velocities, and the initial blower torque setting.
Page 29
(directly exhausted) cabinet that could operate at a constant exhaust system vacuum of only 0.25 inches of H O. Realizing that this was impossible for existing Type B technologies, Labconco developed a new type of BSC; The proposed Type C1. Figure 1-8...
Page 30
Unique to the Axiom is the Chem-Zone, a directly exhausted work zone. The central portion of the work surface is surrounded by grilles on the sides, front and back. Air entering these grilles is drawn to the Exhaust HEPA filter, and when the cabinet is connected to an exhaust system, out of the laboratory.
Page 31
Inlet Relief Valve Also unique to the Axiom is the Inlet Relief Valve, The intake valve is angled such that the incoming make-up (room) air must keep the valve open to maintain normal operation; a loss of exhaust volume will allow gravity and air pressure to close the inlet valve and open the magnetic switch, resulting in an exhaust alarm.
Page 32
Exhaust System Alarm The Axiom is unique in its response to an exhaust alarm. Instead of shutting off the BSC’s blowers, the Axiom starts its “Activate Protection Protocol” – a certifier- programmed interval of up to 5 minutes. During this interval, the Axiom’s exhaust blower speeds up to compensate for the loss of the remote exhaust fan to provide constant inflow.
Page 33
Figure 1-8d Inlet Relief Valve Open Inlet Relief Valve Closed Airflow Patterns of the Axiom in Recirculating and Exhausted Modes. ECM Motor Product Service 1-800-522-7658...
Page 34
The modern Class II Biosafety Cabinet was developed in the early 1960’s as a result of the increased availability of High Efficiency Particulate Air (HEPA) filter technology. The motor of choice at the time to drive the cabinet’s blower was the Permanent Split Capacitor (PSC) type.
Page 35
testing a prototype model to a variety of airflows and filter pressures, the motor/blower can be “characterized” for that size cabinet. This characterization allows for the creation of a model-specific motor-blower curve. By programming the curve values into the motor during assembly, the motor will deliver the correct volume of air as the HEPA filters load.
Page 36
UV Relay Light Relay Outlet Relay RS-232 Board Location The Axiom also has an optional contact/RS232 board. This must be installed if the user wants to signal external building monitors or controllers, or wants the Axiom to output operating information to a data logger. The optional board is located in the electronics module.
Page 37
Section Two – Installation Requirements This section offers an overview of site requirements and how to prepare the unit for operation. Prerequisites Before you install the Axiom, you need to prepare the site for installation. Examine the location where you intend to install the cabinet. You must be certain that the area is level and of solid construction.
Page 38
Location Requirements Note: The biosafety cabinet should be located away from traffic patterns, doors, fans, ventilation registers, fume hoods and any other air-handling devices that could disrupt its airflow patterns. All windows in the room should remain closed. Figure 2-1 shows the preferred location for the biosafety cabinet. Figure 2-1 Preferred location Air register blocked or redirected...
Page 39
CIRCUIT BOARD THAT HAS DRY CONTACTS THAT CAN BE CONFIGURED TO SIGNAL THE EXHAUST BLOWER TO TURN ON WHEN THE AXIOM’s BLOWER IS ON. FOR FURTHER INFORMATION, CONTACT LABCONCO’S PRODUCT SERVICE DEPARTMENT. AS AN ALTERNATIVE, A REMOTE ELECTRICAL SWITCH FOR THE EXHAUST BLOWER CAN BE INSTALLED NEAR THE AXIOM.
Page 40
Figure 2-1c The Inlet Relief Valve located on the top of the cabinet is designed to draw a maximum of 100 CFM (2.83 m /m). Attempting to draw additional room air through the valve (room air exhaust), can result in unstable cabinet operation.
Page 41
The exhaust system must be capable of moving the following volumes of exhaust air at the negative pressures listed. The Airflow Volumes are the values recorded via a direct measurement using a flow hood at the cabinet. The Concurrent Balance Values are measured in the exhaust duct via traverse methodology, and will always be higher due to differences in volume measurement methodologies.
Page 42
Note: On 100 and 115 VAC models, both electrical outlets are protected by a ground fault interrupter circuit (GFIC). Labconco does not recommend plugging the biosafety cabinet into a GFIC outlet. Electrical outlets in the cabinet are restricted to 5 amps maximum current.
Page 43
Service Line Requirements All utility service lines should be ¼ inch O.D., brass, copper, or stainless steel, and equipped with an easily accessible shut-off valve. If the service line pressure exceeds 40 PSI, it must be equipped with a pressure regulator to reduce the line pressure.
Page 44
THIS PAGE INTENTIONALLY BLANK Product Service 1-800-522-7658...
Page 45
Section Three – Installation Now that the installation is properly prepared, you are ready to inspect, install, and certify the Axiom biosafety cabinet. This chapter covers how to: • Unpack and move the biosafety cabinet. • Install the cabinet. • Connect the electrical supply source.
Page 46
Note: United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery. Do not return goods without the prior authorization of Labconco. Unauthorized returns will not be accepted. If the cabinet was damaged in transit, you must file a claim directly with the freight carrier.
Page 47
When lifting the cabinet DO NOT lift the cabinet in the middle front area of the hull. Lifting here may bend or distort the bottom of the cabinet, causing damage to the unit. Installing the Biosafety Cabinet on an Existing Work Surface Note: The biosafety cabinet is very top heavy.
Page 48
Installing the Cabinet on a Labconco Base Stand Labconco offers accessory Base Stands in a variety of configurations to suit your particular needs. Stands can be ordered with adjustable telescoping legs or with a manually or electrically adjustable hydraulic lift.
Page 49
Note: If the research involves the use of toxic compounds or volatile materials, contact the facility’s safety officer or Labconco to ensure that the biosafety cabinet and its exhaust system are compatible with the materials you will be working with.
Page 50
Configuring for an Exhaust System Connection Note: The Type C1 is configured at the factory to operate in recirculating mode (NOT connected to an exhaust system). When operating in this mode, the cabinet ignores the inlet relief valve position. If you intend to connect the Type C1 to an exhaust system, YOU MUST RECONFIGURE IT FOR THIS INSTALLATION.
Page 51
b) Electronic configuration blue bold italic Keypad operations are shown as Menu screen selections are shown as green italics Menu 1. With the unit in operation, access the menu by pressing the button. The Service display panel will show the first level menu. Select menu by pressing the Service buttons until the...
Page 52
The interval can be programmed from 0-300 seconds. Note: Consult with your facility’s safety officer or Labconco to help establish how long the Axiom should continue to operate after an exhaust system failure. 6. The third screen allows you to set the sash height at either 8 or 10 inches.
Page 53
7. The fourth screen allows you to select whether the Axiom has a UV light or not. IT IS CRITICAL THAT THE YOU LEAVE THE UV LIGHT CONFIGURATION AS IT WAS SET AT THE FACTORY. Press to return to the first Configuration screen. The unit is now properly configured for operation connected to an exhaust system.
Page 54
Disconnecting the Axiom from an Exhaust System Connection to Operate in Reciculation a) Mechanical configuration 1. Unplug the Axiom. Using an appropriate ladder or platform, remove the exhaust cover panel on the top left side of the exhaust cover, and install it on the studs on the right side.
Page 55
b) Electronic configuration blue bold italic Keypad operations are shown as Menu screen selections are shown as green italics Menu 1. With the unit in operation, access the menu by pressing the button. The Service display panel will show the first level menu. Select menu by pressing the Service buttons until the...
Page 56
3. The screen will now prompt you for the password; it is: Light UV Light Timer Timer Note: Any other key sequence will return you to the Attention screen. 4. The first screen allows you to select whether the Axiom is connected to an exhaust system or not. IT THIS CABINET HAS IS CRITICAL THAT THE NO EXHAUST CONNECTION...
Page 57
Connecting the Biosafety Cabinet to Utility Service Lines Note: Some models have a solenoid valve connected to the service valve on the right side, rear position. The solenoid prevents gas from flowing to the service valve when the unit blower is off. It is the only service valve position that can be fitted with a solenoid valve.
Page 58
Optional Vacu-Pass Cord & Cable Portal Use Note: There must be enough clearance to pass the cord or cable between the Axiom’s exterior dress panel and any obstruction. Note: Some Vacu-Pass components and the cord or cable passing through it may become contaminated during use of the cabinet.
Page 59
Drain Valve Installation In order to prevent damage during shipping, the drain valve assembly has not been installed. If desired, the valve should be installed after the cabinet is in its final location. To install the valve assembly, follow these steps: 1.
Page 60
THIS PAGE INTENTIONALLY BLANK Product Service 1-800-522-7658...
Page 61
Section Four - Initial Operation of the Axiom This section will take you through the steps to begin initial operation of the Axiom, with an explanation the controls, display and options, in order to prepare the unit for its initial certification. System Reset Switch The biosafety cabinet has a system reset switch for resetting its microprocessors.
Page 62
Information Center The Information Center is an LCD display located on the right side wall at eye level. When the blower is started, if the Axiom is configured to be connected to an exhaust system, Figure 4-2a will appear for 60 seconds while the unit initiates operation. If the unit is not connected to an exhaust system, screen 4-2b will appear for the first 60 seconds of operation.
Page 63
Figure 4-4a Alarm Screens Sash is too high The sash is open too far for safe operation. Figure 4-4b Airflow Alerts The airflow patterns in the cabinet have changed, resulting in a sudden change in either motor speed. This is most likely due to a blockage of the grille or the exhaust filter outlet.
Page 64
Operating the Sliding Sash The counterbalanced, anti-racking sash mechanism requires only a few pounds of force to move the sash up or down. You can open or close the sash smoothly with one or two hands positioned on either handle. The sash position alarm and safety interlock system senses the sash position and acts appropriately.
Page 65
The Axiom Touchpad The touchpad of the Axiom is shown in Figure 4-6. Take a moment to get familiar with the buttons, their locations and functions. Also familiarize yourself with the display located on the right side wall. The display will report system functions, such as filter capacity, timer displays, alarm or error messages, as well as icons that illuminate when cabinet functions such as UV light and blower are operational.
Page 66
Navigating the Axiom Menu Screens MyLogic allows you to use the Smart-Start or Night-Smart features that activate functions automatically when the sash is opened or closed. Night-Smart will only work if the cabinet is not connected to an exhaust system. If equipped, the UV lamp can be programmed to operate for a given time interval when the sash is closed, before it shuts off.
Page 67
Navigating the MyLogic Menu Screens The MyLogic screens will allow you to set the cabinet’s clock, and to personalize its operation. Please note all MyLogic screens have a blue background. Setting the Clock MyLogic Using the buttons on the touchpad, highlight the option-it will turn ▲...
Page 68
Configuring the Axiom In the first MyLogic screen, use the ▲ ▼ configure my buttons on the touchpad to highlight Logic for use -it will turn white when selected. Press OK/MUTE to enter the first configuration screen: The first screen gives you the option of activating the Smart-Start option for the blower;...
Page 69
If you choose to use NightSmart option for the UV lamp, this screen allows you to control the time the UV lamp will remain on after the sash is closed. Use the ▲ ▼ buttons on the touchpad to cycle through the time OK/MUTE intervals available, and then press to select it.
Page 70
Startup tone Using the buttons on the touchpad, highlight the ▲ ▼ Startup Tone option-it will turn white when selected, and then press . Using the buttons on the ▲ ▼ Turned on Turned off touchpad, highlight either option. When turned on, an audible beep will sound during the first 60 seconds of blower operation to caution the user that the unit is not yet ready for use.
Page 71
UV Settings For models equipped with the optional UV light the Axiom has an integral UV light maintenance system. It allows you to define how many hours you want the UV lamp to operate before receiving a reminder to replace it, a way to monitor how many hours the lamp has been on, and the means to reset the UV lamp hourmeter.
Page 72
Timer Operation NOTE: The timer button allows activation of an interval (countdown) or stopwatch (elapsed) timer. The timers cannot be operated simultaneously. Timer To access the main timer menu, press the button anytime during normal operation. The main timer menu is shown on the LCD display. Use the buttons ▲...
Page 73
NOTE: In order to properly certify the Axiom, you must obtain an Axiom Certifier Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by contacting Labconco Customer Service Department at 800-821-5525. Note: The ECM motor is programmed by Labconco to deliver a constant volume of air throughout the life of the HEPA filters.
Page 74
Light UV light 8. When requested for the password press Timer Timer OK/Mute then 9. If the password is properly entered, the first certification screen will display the supply blower motor’s PWM setting. Press the ▲ ▼ button to increase or decrease the blower speed as OK/Mute needed.
Page 75
Axiom Filter Life Gauge Operation and Resetting for Local Conditions The filter life gauge calculates the remaining filter life by measuring and comparing the current motor speed to the initial speed noted at the factory during assembly, and the calculated maximum motor speed, using the formula: 100-(100x (current speed-original speed) ) = % filter life left max.
Page 76
HEPA Leak Test Information All Axiom models were tested to the HEPA Filter Leak Test as described in ANSI/NSF Standard 49:2012. Note: ensure all internal surfaces of the Axiom have been appropriately decontaminated before proceeding. Supply HEPA Filter Leak Test All Axiom models were tested to the HEPA Filter Leak Test as described in ANSI/NSF Standard 49:2012.
Page 77
If the Axiom has been used with biohazardous, toxic chemicals, or radioisotopes, the DOP access line cannot be used. The upstream concentration cannot be measured directly. Testing at Labconco has shown the actual concentration varies from the calculated value. Table 5-1 shows the actual vs. calculated concentrations for the Axiom models.
Page 78
3. Pull the tube toward the front of the unit. Remove the access tube cover. 4. Using an appropriate photometer, connect the unit's upstream line to the access tube. Note: Keep the photometer sampling tube as short as possible. It should never exceed 6 feet maximum.
Page 79
2. The Axiom is now ready for the Supply HEPA Filter Leak Test. Set the photometer sampling valve to “DOWNSTREAM”. Ensure there is a proper vacuum at the sampling port of the pistol. 3. Open one Laskin nozzle. 4. Ensure the pressure gauge on the generator is reading 22 +/- 1 PSIG. 5.
Page 80
Figure 5-7. Figure 5-7 Testing at Labconco has shown the actual concentration varies from the calculated value. Table 5-2 shows the actual vs. calculated concentrations for the models. Table 5-2...
Page 81
4. If connected to an exhaust system, secure the Inlet Relief Valve in an open position, to keep the cabinet blowers operating, and then remove the screws that secure the front of the exhaust connection to its structure. Place the front panel aside.
Page 82
Downflow Velocities All Axiom models, are classified as having uniform downflow, and the average downflow velocity for all 4-foot models is 65+/-5 FPM, and all 6-foot models is 55+/-5 FPM, tested as per ANSI/NSF Standard 49:2012. All readings should be in a plane 4 inches above the bottom edge of the sash, and the UV Lamp, IV bar and all other accessories must be removed before performing downflow test.
Page 83
Figure 5-8 Figure 5-9 Acceptance Average downflow shall be 65 +/-5 FPM for all 4-foot and 10” 6-foot models, 55 +/-5 FPM for 8” 6-foot models, with all test points being within +/- 16 FPM of the overall average. Product Service 1-800-522-7658...
Page 84
Inflow Velocity Calculation - Primary Method Note: The dress panel MUST be installed on the cabinet to obtain an accurate inflow value. The average inflow velocity for Axiom models is 105 ±5 FPM. It should be determined by converting the inflow volume to the average inflow velocity. The inflow volume is measured directly by sealing a direct inflow measuring flow meter to the face of the unit.
Page 85
NOTE: You must use Labconco Certifier Kit, Labconco P/N 3858403 to perform this test properly. Qualified certifiers can obtain this kit by contacting Labconco Customer Service Department at 800-821-5525.
Page 86
Figure 5-11 The Sensor Distance, as described in Table 5-5, using the slots cut in the 6-foot Logic template. 4.00-inch mark 3.25-inch mark Figure 5-12 The Sensor Distance, as described in Table 5-5, using a scale. Table 5-5 Secondary Inflow Test Specifications – 8 Inch Sash Opening Cabinet Width (feet) Sash Opening Template Sensor distance (inches)
Page 87
Locate the single row of holes at the front of the grille. Mark the 6 hole from the side wall and subsequent test points every 9 holes until the number of test points marked equals the width of the cabinet in feet (for a 3-foot cabinet, for example, mark the first 3 points).
Page 88
Concurrent Balance Values Axiom Inflow and Inlet Valve flow numbers are established by a direct measurement of air flow. Traditional exhaust system measurements are made using either an anemometer or Pitot traverse methodology. Because of these different methodologies, the total volume of air flowing through the exhaust system will vary. Table 5-6 Concurrent Balance Volume Specifications Model...
Page 89
Work Area Air Cleanliness Test (optional) If the customer wants to verify air cleanliness in the work area, use the following procedure: 1. Mark test points at the geometric center of the work surface, and midway between the center test point and each side wall. 2.
Page 90
Fluorescent Light Level (optional) The fluorescent light intensity should be measured as described in ANSI/NSF Standard 49:2012, Annex F. NOTE: The light meter should be color- and cosign corrected for accurate readings. 1. Test points should be marked along the centerline of the work surface, starting 6 inches from the side wall, with subsequent points 12 inches apart until the center of the work surface is reached.
Page 91
Vibration (optional) The vibration level should be measured as described in ANSI/NSF Standard 49:2012. Note: The vibration meter should have a range of 20-20,000 Hz, and have an output in displacement. 1. Vibration level should be recorded from a single point, on the geometric center of the work surface.
Page 92
Ground Fault Circuit Interrupter Test (115/100 Volt models only) Note: The GFCI tester must be able to simulate a fault of 3mA. 1. Place the tester into any of the Axiom’s outlets. The indicator lights of the tester should indicate normal function, and the GFCI’s indicator LED should be lit. 2.
Page 93
This section will review the tools needed, and common service operations for the Axiom. Recommended Tools for Service The tools needed for most common service operations are: ✓ Certifier Kit, Labconco Part # 3858403 ✓ #2 Phillips Screwdriver or long bit for cordless drill ✓ ½-inch Socket with extension and ratchet ✓...
Page 94
External Dress Components Front Dress Panel Removal and Installation Figure 6-1 1. Locate and remove the two Phillips screws that secure the front panel as shown in Figure 6-1. They are located on the bottom corners of the front dress panel Swing the bottom of the dress panel out to clear the fluorescent light and then lift the front dress panel straight up and away from the cabinet.
Page 95
Side Panel Removal Note: The side panel(s) must be taken off to remove the Axiom from its pallet. 1. Remove and save the two Phillips screws on the front edge of the side panel. 2. Pivot the front of the panel away from the cabinet. Lift the panel straight up and pull the panel straight away from the cabinet to release the rear of the side panel from its hinge.
Page 96
If the corner post lower trim pieces do not fit flush to the bottom edge of the metal corner post, examine the front flange of the chassis. The flange should be straight. If it is bent back, it can be straightened by bending the flange forward with an adjustable wrench, as shown in Figure 6-4.
Page 97
Sash Operation The Axiom sash uses a deadweight-counterbalanced system. The system consists of two separate weights on either side of the liner. Both weights are synchronized by a locked axle-pulley system, so that both weights must move together, preventing the sash from racking. Adjustment There are no adjustments for the width of the sash track.
Page 98
Sash Position Switches Four microswitches are used to indicate the sash position. The switches are the same, Labconco part # 3832400. A photo of the switches is shown in Figure 6-6. Figure 6-6 Sash Closure Position Switch Sash Restricted Opening Switch 8”...
Page 99
8”- and 10” Sash Operating Position Switches The sash operating position switches signals the display microprocessor that the sash is open too far for proper operation, as shown in Figure 6-8. Both switches are wired in the normally open configuration, and the display microprocessor only responds to the appropriate switch selected during configuration.
Page 100
Sash Cable Retainer The sash cable retainers are located at the top of each sash pulley, as shown in Figure 6-9. If the retainer needs adjustment, Use a 3/8-inch wrench to loosen the bolt securing the retainer and adjust it until it there is approximately 1/16” gap between the retainer and the pulley.
Page 101
Figure 6-10 Exhaust filter clamp bolts (two in each side of the exhaust connection) To Install an Exhaust HEPA Filter: 1. Slide the new exhaust HEPA filter all the way into the pan. 2. Sequentially tighten the four exhaust filter sealing bolts located at each corner of the exhaust filter tray until the exhaust filter gasket is compressed no more than 50%.
Page 102
NOTE: During disassembly, the blower plenum cover can rest on the center sash support, as shown in Figure 6-11, to allow repositioning of the panel or your grip on it. Figure 6-11 Center Sash Support 3. Use 1/2-inch sockets or wrenches to tighten the three plenum lift bolts located as noted in Figure 6-12.
Page 103
To Install a Supply HEPA Filter: 1. Install a new supply HEPA filter by pushing it straight into the cabinet, ensuring that it is correctly oriented, and that the filter is centered on the filter shelf. 2. Use the 1/2-inch sockets or wrenches to sequentially loosen the three plenum lift bolts located as noted in Figure 6-12.
Page 104
Figure 6-13 ECM Power Connector Release Tabs ECM Data Connector Release Tab 6. Use a Phillips screwdriver to remove the two sealing plate screws and the sealing plate. Using a 1/2-inch wrench, remove the three blower mounting bolts, as shown in Figure 6-14. Figure 6-14 Blower Mounting Bolts (3) Sealing plate screws (2)
Page 105
8. To replace the motor/blower assembly, reverse the above procedure. Note: During reinstallation of the blower assembly, push the blower into position, and loosely install the two blower mounting bolts. Press the sealing plate against the side of the blower to ensure it is properly sealed against the back of the blower mount.
Page 106
filter and set it in a safe spot. 7. Disconnect the exhaust motor harness at the pass through near the top of the chamber, as shown in Figure 6-17. Figure 6-17 Exhaust Motor wiring harness connectors NOTE: It will ease plenum installation and removal by laying two 2x4 or 4x4 supports under the plenum, and over the filter chamber flange, as shown in Figure 6-18.
Page 107
Center Work Surface Removal Note: The work surface must be thoroughly decontaminated before removing it from the cabinet. Lift either of the wing work surfaces up by grasping the knob handles at either front corner. This will lift up the center work surface. Lift the front edge up, pivoting the work surface on its back edge until the front edge of the work surface engages the catch located on the back wall of the work area.
Page 108
Figure 6-19a Wing work surfaces Center Work surface Product Service 1-800-522-7658...
Page 109
Grille Removal Note: The grille must be thoroughly decontaminated before removing it. 1. Remove the work surface as described earlier. 2. At one end of the grille, grip the front of grille with one hand, and the back with the other hand. Pivot that end of the grille upward and inward, paralleling the angle of the sash, as shown in Figure 6-20.
Page 110
Diffuser Removal Caution: The diffuser should be surface decontaminated before removal. Note: If an IV bar is in place, it will have to be removed first. 1. Raise the sash to its fully open position. 2. Support the front edge of the diffuser while removing the two 7/16-inch acorn nuts on the front of the diffuser, as shown in Figure 6-21.
Page 111
4. To reinstall the diffuser, ensure that the two holes on the back edge of the diffuser align with the diffuser support pins on the rear wall of the work area, as shown in Figure 6-23. Figure 6-23 Diffuser pins engaging the holes in the rear of the diffuser.
Page 112
Preparing the Axiom for Gaseous Sterilization Note: This section only reviews the steps required for preparing the Axiom for Gas Sterilization. Thoroughly understand the sterilization procedures and protocols supplied by the manufacturer of the sterilizing system before attempting this operation. Unplug the Axiom from electrical power.
Page 113
Section Seven – Electrical System Service Operations NOTE: In order to properly service the Axiom, you must obtain an Axiom Certifier Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by contacting Labconco Customer Service Department at 800-821-5525. Fluorescent Lamp Lamp Identification All Axioms use dual fluorescent lamps.
Page 114
30 = the lamp Wattage T8 = the lamp mounting configuration UV lamp Specifications • For all models, the UV lamp number is G30T8. The Labconco part number is 1271300. • For all 115/230 VAC models, the ballast assembly is Labconco part number 3829901.
Page 115
Changing the UV Lamp Note: For optimum performance, the UV lamp should be changed on an annual basis. Caution: The UV lamp and the work area of the cabinet must be thoroughly surface decontaminated before removing the lamp. 1. Unplug the Axiom. 2.
Page 116
Electronics Module Access The Electronics module is located on the top right side of the cabinet. To access the module, do the following: 1. Unplug the Axiom, either at the wall, or at the power cord inlet at the back of the electronics module, as shown in Figure 7-2.
Page 119
Power Supply Board Service Operations Fuse Testing & Replacement The Power Supply Board is protected by an integral 5 x 20mm, fuse, rated at 2.5 Amps at 250 volts, as shown in Figure 7-5. The fuse is tested with a standard continuity meter. Any resistance across the fuse indicates a defective fuse and should be replaced.
Page 120
Display Board Service Operations Display Board Diagnostics is addressed in the section on Touchpad Test Operations. Touchpad Connections To correctly connect the touchpad to the Display Board, orient the ribbon connector so the metallic side of the connector is away from the side of the Axiom, as shown in Figure 7-6.
Page 121
Use caution when testing electrical components. NOTE: To properly measure motor line and signal voltages, you should use the pass-thru harness, Labconco Part # 3842601. The harness is part of Labconco Certifier’s Kit, Part #3858400. Qualified certifiers can obtain this kit by contacting Labconco Customer Service Department at 800-821-5525.
Page 122
Figure 7-7 15 pin connector 4. Disconnect the 15-pin connector, and install the pass-thru harness, as shown in Figure 7-8. Plug the Axiom back in. With the meter set on AC voltage, verify the AC mains line voltage is present within ± 5 VAC when the AC voltage is measured across pin 2 (black wire) and pin 3 (white wire) on the 3-pin connector, as shown in Figure 7-9.
Page 123
6. Next, connect the meters + DC voltage lead to pin 6 (yellow wire) and verify there is 12 Volt DC, as shown in Figure 7-10. The presence of 12 Volt DC on this yellow wire tells the motor to start. Figure 7-10 Measuring 24 VDC voltage 7.
Page 124
Motor Output Signal Testing Use the Service Menu’s “Current Settings” screen to access the Supply and Exhaust Motors’ Speed Factor, Motor RPM, average RPM, and new filter RPM settings. Figure 7-11 shows the LCD screen and the location of these values. Figure 7-11 Motor Speed Factor Setting Motor Speed in RPMs/...
Page 125
Wiring Diagrams -115 VAC Models Product Service 1-800-522-7658...
Page 126
Wiring Diagrams -230 VAC Models Product Service 1-800-522-7658...
Page 127
Section Eight – Purifier Axiom Accessories Air Flow Sensor Kit No. 3405500 Note: A qualified certifier must install and calibrate this sensor to obtain accurate operation. Note: Exercise caution when working on the exhaust filter cover, to prevent damaging the Exhaust HEPA filter. Installation Procedure-Units without the Ventus Canopy Connection 1.
Page 128
Installation Procedure- Units with the Ventus Canopy Connection Axiom Type C1 cabinets Remove the two screws that hold the airflow sensor into its body, as shown. DO NOT pull the sensor out of the body. If you are installing the canopy airflow sensor into a canopy connection, install the bracket onto the airflow sensor as shown.
Page 129
5. For the Axiom, route the sensor cable from the sensor, through the grommet, and outside of the exhaust duct. Reinstall the grommet. Ensure the cable does not hinder the operation of the front relief valve. 6. Note the orientation of the sensor harness plug and the sensor;...
Page 130
Calibration Procedure- NOTE: YOU MUST FOLLOW THESE INSTRUCTIONS EXACTLY, OR THE SENSOR WILL NOT OPERATE PROPERLY! Note: You will need to access the certifiers menu to calibrate the airflow sensor. Note: To properly calibrate the air flow sensor, a qualified certifier must verify the inflow and downflow values.
Page 131
Light 15. When requested for the password press light Timer Outlets OK/Mute then . The cabinet blower(s) will stop. 16. If the password is properly entered, the first certification screen will display the blower motor’s PWM setting. 17. The unit must sit for approximately 60 seconds with no airflow over the sensor in order to establish the zero point value.
Page 132
Confirming Calibration Procedure- With the cabinet operating: MAIN ME NU Menu 1. Press the button. Services 2. Press the button until the option is highlighted 1 . MyL ogic ▼ (it will turn yellow). 2. S et t ings OK/Mute 3.
Page 133
Additional service fixtures can be installed in any of the four available service fixture holes in the sidewalls. The service fixture kit is Labconco part number 3747500. Note: Some models have a solenoid valve connected to the service valve on the right side, rear position.
Page 134
Tube fitting Tube fitting nut 10. Make sure the service fitting in the Purifier is in the off position before pressurizing the service line. Once pressurized, check the valve body, the coupling, and the tube fitting for leaks. Tighten any leaking joints. 11.
Page 135
UV Lamp Kit No. 3858500, -01, -02 All Labconco Purifier Axiom Series Biosafety Cabinets are pre-wired for the installation of a UV lamp kit, consisting of the UV lamp, its ballast, and fasteners. The kit is part number 3858500 for 115 Volt models, 3858501 for 100 Volt models, and 3858502 for 230 Volt units.
Page 136
5. Using the screws, flat washers and locking nuts provided, secure the UV ballast to the top of the electronics module. 6. Wire the UV ballast into the wiring harness, following the wiring diagrams located in Figure 2a-c. 7. Reattach the electronics module to the top of the cabinet before reconnecting the electrical power.
Page 137
The interval can be programmed from 0-300 seconds. Note: Consult with your facility’s safety officer or Labconco to help establish how long the Axiom should continue to operate after an exhaust system failure. 11. The third screen allows you to set the sash height at either 8 or 10 inches.
Page 138
IV Bar Kit No. 3858600, -01, -02, -03 Completely decontaminate all interior work surfaces of the Purifier before beginning the installation of the IV bar. Installation Procedure 1. Disconnect power to the Axiom and raise the sash to the fully open position. 2.
Page 139
Vacu-Pass Cord & Cable Portal Installation Note: You will need to access the right side of the cabinet in order to install the Portal Note: There must be enough clearance to pass the cord or cable between the exterior dress panel and any obstruction. Note: Some Vacu-Pass components and the cord or cable passing through it may become contaminated during use of the cabinet.
Page 140
RS-232 Connection Kits RS-232 Kit 3858310 (9 Pin), 3858311 (25 Pin) The kit contains the RS-232 PCB (P/N 3402200), the attaching hardware and connecting cables to install the PCB inside the electrical control box (located on top of the Axiom). Picture 1 shows an electrical control box with the Axiom RS-232 PCB installed.
Page 141
2) Locate the black square connecter labeled as J2 and the white connector labeled as J3 on the RS-232 PCB, see picture 3. These two connectors show through the electrical control box as shown in picture 1. Connector J2 is the RS-232 PORT connector.
Page 142
4) Locate the internal RS-232 cable (P/N 3432000) included in the kit and connect it between the RS-232 PCB J1 and the Transition PCB (P/N 3431500) J6 as shown in picture 5. Picture 5 5) Re-secure the electrical control box to the top of the Axiom. 6) Connect the computer interface cable (P/N 7537800 –...
Page 143
1. Computers with a 25-pin D-sub male serial connector should use Connect Cable, Labconco part number 7537801, to connect a computer to the Axiom. 2. Computers with a 9-pin D-sub male serial connector should use Connect Cable, Labconco part number 7537800, to connect a computer to the Axiom.
Page 144
OPTIONAL AXIOM RS-232 DATA OUTPUT Here is the RS-232 output for the Axiom. As is standard for all Labconco units, the data is limited to 80 characters, including spaces, etc. Operating Status: Columns 1-4: Column 1: UV Light --- U=on, O=off...
Page 145
Exhaust Blower PWM Duty Cycle, Columns 49-50: Blower speed ---- three digit 0-100% indicates the duty cycle of the motor speed signal, followed by a comma Exhaust Instantaneous Blower RPM, Columns 51 -55 Most recent blower RPM read in from motor, 4 digits, updated every second, followed by a comma Exhaust Average Blower RPM, Columns 56 -60 An averaged value, 4 digits, from the last 16 blower RPM readings above;...
Page 146
4. The “Connect To” dialogue box will open. Using the down arrow selection button, select the communication port to which the cable has been connected. Press “OK.” 5. The “Com X Properties” dialogue box will open. Enter the appropriate data properties and press “OK.”...
Page 147
Labconco unit is turned on. With the optional RS-232 board installed, the Labconco Axiom BSC provides two contacts: One to indicate the internal Labconco blower is on, (so that external blowers may also be energized) and the other to indicate airflow errors. These contacts provided will “short circuit”...
Page 148
The generalized circuit shown below indicates how to interface the Labconco contacts to power a high power device. Note that the Labconco contacts are used to power the low-power coil of an added external power style relay/contactor—in this case the relay is a Hasco HAT903ASDC12, and the 12V coil draws only .08Amps (the Labconco +12V is...
Page 149
AC mains rated coil voltage is shown below – it does not utilize the FUSED+12V and CIRCUIT COMMON from the Labconco Axiom. Note: A W199 Magnacraft Series relay can switch up to 300Vac @ 40 Amps……….it’s coil is rated for 12 Vdc, 171mA, or approx.
Page 150
The final diagram shows a connection for a 3-phase powered blower. In this case, a 12 VDC relay coil is used, and it gets its power from the Axiom. Product Service 1-800-522-7658...
Page 151
This set of contacts (dry contacts pins 1 & 2 of J3 on the electrical enclosure, and the ORANGE wires on Labconco harness P/N 3432300) will indicate when the Axiom has detected an airflow error, as described by the conditions below:...
Page 152
Section Nine – Miscellaneous Troubleshooting Refer to the following table if the Biosafety Cabinet fails to operate properly. If the suggested corrective actions do not solve the problem, contact Labconco for additional assistance. PROBLEM CAUSE CORRECTIVE ACTION Cabinet blower and...
Page 154
PROBLEM CAUSE CORRECTIVE ACTION Blower runs Motor signal wires Measure motor output signal slower than damaged or voltage. Inspect signal wiring and normal; speed disconnected connections control has little effect on blower speed. Blower may take more than 5 seconds to start. Power Supply Test output voltages on the Power Board...
Page 155
PROBLEM CAUSE CORRECTIVE ACTION Fluorescent light Lamp(s) are Replace defective lamp(s) is dim or defective flickering Lamp wiring is Inspect lamp wiring. disconnected Defective lamp Replace lamp ballast. ballast UV light not Sash is open Close sash – UV light will not work working with the sash open.
Page 156
exhaust outlet not blocked or restricted. Blockage or Ensure that the towel catch and restriction under plenum beneath the work surface the work surface are unobstructed. Contamination of Improper technique See “Use of the Cabinet” section in work in the or procedure for the manual.
Page 158
Dimensions – 4 Foot Models Product Service 1-800-522-7658...
Page 159
Dimensions – 6 Foot Models Product Service 1-800-522-7658...
Page 160
Section Ten – Vacuum Start Switch (VSS) Background The VSS is a single pole vacuum switch, added to all Axiom models with serial numbers higher than 1709xxxxx. It is attached to the exhaust connection of the Axiom, as shown in Figure 10-1a and b. Figure 10-1a &...
Page 161
The operational logic of the VSS is shown in Figure 10-2. Figure 10-2 VSS Diagnostic Screens Figure 10-3 If the VSS sees insufficient vacuum at startup, the VSS Airflow check screen will be shown when the “Blower” button is pressed, as shown in Figure 10-3.
Page 162
Figure 10-4 If, after 30 seconds, the VSS still sees insufficient vacuum, the exhaust alarm screen is activated, and the alarm locks out, requiring the blower button be pressed again to attempt a blower restart. The screen is shown in Figure 10-4. If there is no red “V”...
Page 163
VSS Calibration VSS is calibrated during production. If recalibration is required, however, follow these steps: Figure 10-5 1) Access the Calibration menu. The first screen displayed after entering the password will be the menu selecting whether to adjust airflows and filter life gauges, or adjust the VSS.
Page 164
Figure 10-7 3) The second VSS calibration screen will stop the Axiom’s blowers. It is critical you allow the Axiom 10-15 seconds for all blower movement to stop; this will be required for proper calibration. When the VSS is closed, the Axiom’s speaker will repeatedly beep.
Need help?
Do you have a question about the Purifier Axiom 30441 and is the answer not in the manual?
Questions and answers