Figure 6-17
Look to identify the cause of the airflow disruption, and remove the blockage or return
the work surface to its proper position.
When a blockage of the airflow occurs, the cabinet's blower automatically increases its
speed to maintain constant volume airflow. This is a protective feature; however, if the
disruption is significant (for example blocking the entire front grille) it will not guarantee
product or personnel protection remains during the significant blockage event.

Resetting the Airflow Alert System

The Airflow Alert automatically dismisses once the motor speed has stabilized.
Note: Once the blockage or disruption has been resolved, the Airflow Alert may
reactivate while the blower returns to correct operating speed. It will dismiss
automatically once the blower reaches correct operating speed.

Alarms

Any alarm that activates requires the user's immediate attention, and some form of
actionable response to clear the alarm.