User-Supplied Circulator Requirements – Temp-Zone™ Package Only
The Temp Zone™ features a temperature controlled area on the work surface where
heated or chilled samples may be transferred and kept at optimum temperature. A user-
supplied circulator that continually delivers heated or chilled water at a minimum flow
rate of 0.5 GPM (1.9 LPM) with typical volume of 6 liters is required and may be
obtained from your laboratory supply dealer.
Quick connection fittings are included on the left side of the cabinet and are sized for
3/8" diameter polyethylene tubing. As a convenience, the following items are included
with the Temp-Zone package:
• 20 Feet (6 meters) of 3/8" (9.5 mm) outer diameter tubing
• Insulation for 3/8" outer diameter tubing
• Quick-connect fittings (3/8 tube x 1/8 NPT, 3/8 tube x 1/4 NPT, 3/8 tube x 3/8
NPT)
The above items will allow for connection to a water circulator. The water circulator can
be located underneath the cabinet or next to the cabinet on the floor.