Sash Operation; Blower Operation; Light Operation; Outlet Operation - Labconco Cell Logic+ B2 User Manual

Biosafety cabinets
Hide thumbs Also See for Cell Logic+ B2:
Table of Contents

Advertisement

Sash Operation

The counterbalanced, anti-racking sash mechanism requires only a few pounds of force
to move the sash up or down. You can open or close the sash smoothly with one or two
hands positioned on either handle.
The sash position alarm and safety interlock system senses the sash position and acts
appropriately. The cabinet has been programmed to operate at an 8-inch (203 mm)
sash opening. Raising the sash above or below its operating height will activate the
audible and visual alarms. The audible alarm can be temporarily muted for
approximately five minutes by pressing the
[OK/MUTE]
button on the Keypad. Moving
the sash back to its operating position will reset the alarm. The safety interlock system
senses when the sash is closed and allows the optional UV lamp to operate only when
the sash is closed, to protect the operator from irradiation.

Blower Operation

The Cell Logic+ Type B2 biosafety cabinet can be configured to automatically turn on
the internal blower. If this feature is enabled, raising the sash from closed to operating
height will automatically turn on the blower. If this feature is not enabled, raise the sash
to operating height, and press [BLOWER]. Either method will start the blower, and the
Blower Starting
screen will appear for approximately 1 minute.
If the facility's remote blower, or exhaust system, connected to the cabinet is not pulling
the required volume of exhaust air through the exhaust housing, the internal blower will
not start. If the building exhaust system does not reach the required minimum volume of
airflow within 60 seconds of pressing [BLOWER], an exhaust alarm will occur.

Light Operation

The Cell Logic+ Type B2 biosafety cabinet provides dual LED lamps for illuminating the
work surface. The cabinet can be configured to automatically turn on the LED lighting.
If this feature is enabled, raising the sash from closed to operating height will
automatically turn on the LED lighting. If this feature is not enabled, raise the sash to
operating height, and press [LIGHT]. Either method will illuminate the LED lighting.

Outlet Operation

The Cell Logic+ Type B2 biosafety cabinet provides two GFIC outlets with a maximum
total amperage draw of 5A (100-115v models) or 3A (230v models). Power to the
outlets can be activated or deactivated from the Keypad. To activate or deactivate
power to the outlets, press [OUTLET].
57