If An Airflow Alert Activates; Resetting The Airflow Alert System; Working In The Cabinet - Labconco Logic+ 30348 Series User Manual

Type b2 biosafety cabinets
Hide thumbs Also See for Logic+ 30348 Series:
Table of Contents

Advertisement

If An Airflow Alert Activates

The most common causes of an Airflow Alert are:
 Blockage of the inlet grilles or exhaust outlet
 Removal of the work surface or grille during
operation

Resetting the Airflow Alert System

The Airflow Alert automatically resets to normal operation
once the motor speed has stabilized.

Working In the Cabinet

A more thorough review of using a biosafety cabinet can be found in: Biosafety
in Microbiological and Biomedical Laboratories (BMBL ), published by the
Centers for Disease Control and Prevention (cdc.gov/biosafety/publications).
Planning
 Thoroughly understand procedures and equipment required before
beginning work
 Arrange for minimal disruptions, such as room traffic or entry into the room
while the cabinet is in use
Start-up
 Turn off UV light (if included on your cabinet)
 Slowly raise the sash until the bottom of the sash aligns with the sash
indicator decal located on the left side of the work area (Fig. 5-5)
 Turn on the light and cabinet blower if the automatic features have not been
enabled
 Check the air grilles for obstructions
 Allow the cabinet to operate until the Home Screen is shown
BLOWER
STARTING
12:39
Warm up Screen
5: Using The Cabinet
Home Screen
Figure 5-9
44