Config
Zero Schedule
Schedule
Type
Day of
Week
Day of
Month
Hour
Duration
4-10 | 5830 Moisture Analyzer
Zero Schedule
Use this function to schedule automatic analyzer zeroing.
Never
Daily
Weekly
Monthly
Sunday
Monday
Tuesday
Wednesday
Thursday
Friday
Saturday
Enter Value
Enter Value
Enter Value
AMETEK recommends performing a zero every 24 hours for the first
month of operation. After this start-up period, the analyzer will have
dried down sufficiently to increase the zeroing interval to 1 week.
NOTE
AMETEK recommends that the duration of the zero cycle be set to
120 (i.e., 2 hours).
Schedule Type
Select one choice from the list:
•
Never
•
Daily
•
Weekly
•
Monthly
Day of the Week
Select the day of the week from the list.
Day of the Month
Enter the 2-digit value for the day of the month.
Hour
Enter the 2-digit value for the hour of the day using
a 24-hour clock. Zero is the starting hour.
Duration
Enter the zero period value in minutes.