4-10 | 3050 AMS Moisture Analyzer

Zero Schedule

Use this function to schedule automatic analyzer zeroing.
Config
Zero Schedule
Schedule Type
Select one choice from the list:
Never
Daily
Weekly
Monthly
Day of the Week
Select the day of the week from the list.
Day of the Month
Enter the 2-digit value for the day of the month.
Hour
Enter the 2-digit value for the hour of the day using a 24-hour clock. Zero
is the starting hour.
Duration
Enter the zero period value in minutes.
AMETEK recommends performing a zero every 24 hours for the first
month of operation. After this start-up period, the analyzer will have
dried down sufficiently to increase the zeroing interval to 1 week.
AMETEK recommends that the duration of the zero cycle be set to
NOTE
120 (i.e., 2 hours).
Never
Daily
Weekly
Monthly
Schedule
Type
Sunday
Monday
Tuesday
Day of
Wednesday
Week
Thursday
Friday
Saturday
Day of
Enter Value
Month
Enter Value
Hour
Enter Value
Duration