206

Basic Functions

Options on this tab control access to training mode and provide a workspace to create training scenarios.
Training Mode
Allow Instrument to Enter
Training Mode
Training Pathway Cartridges
Allow Expired Cartridges in
Training Pathway
Operator Direct Observation Checklist
Edit Observation Checklist
Update Observation Checklist
Training Scenarios
Add Training Scenario
i-STAT Alinity — System Operations Manual
To deny access to the training path, unselect the check box. If access
to the training path is enabled, choose one of these options:
Require permissions
No permissions required
Select this check box to allow expired cartridges to be used in the
training path.
Select this option to create a checklist of items to display on the
i‑STAT Alinity instrument. Observation checklist is intended to be
used as a location on the instrument where a super user (trainer)
and an operator can see a comprehensive list of the skills needed to
remain competent. The super user (trainer) observes the operator
performing the skills and communicates to the POCC (administrator)
that the operator successfully completed the list.
A text box displays and up to 40 characters can be entered. To save,
click Add Observation. This will display another text box for an
additional item. When all items have been entered, specify an
identifier for the revision in the Checklist Revision text box.
Select this option after all the checklist items have been created and
a Checklist Revision value was supplied. The checklist items and
checklist revision information will display.
Select this option to open a workspace to create training scenarios.
Be aware of the following:
• For the patient information to be displayed on the i‑STAT Alinity
instrument, the PPID feature in Patient Settings must be
enabled.
• For analyte values to display as critical, abnormal, or normal
values, reference and action ranges in Analyte Settings are
required.
Art: 746300-01 Rev. N
This is the default.
Select to allow access to the
training path regardless of
permissions.
Rev. Date: 02-Nov-2022