104
Symptom
Check for Update button is not enabled
Instrument indicates a connection could not be
made to the Server service, OR
Instrument powers off before completing the
software download, OR
Instrument does not power on within 10 minutes
after screen going blank
Repeated attempts to update software via wired
network connection fail
i-STAT Alinity — System Operations Manual
Action to take
Verify the following as described in the
prerequisites, then repeat the software update
procedure:
• The instrument is running software version
OSi05 or later
• Network Services for software update enabled
on instrument
• The battery charge level is no less than 51%
Verify the following as described in the
prerequisites, then repeat the software update
procedure:
• i-STAT Alinity instrument(s) with appropriate
Network Configuration file installed
• Base Station with power cable connected to AC
mains power
• Ethernet Cable connected to healthcare
organization network
• The healthcare organization network is
allowing the instrument to connect to the
internet in order to have access to the Abbott
Managed Server or SDi.
Follow the procedure to perform software update
using USB memory device
Art: 754496-01 Rev. E
Rev. Date: 02-Nov-2022