Mode
NIV
In the table above, A/C mode requires 6 STATNotes items. To check whether these STATNotes items exist:
Click Manage and Assemble Profiles > STATNotes > Manage Items.
The screen displays all existing items.
If all of the STATNotes items needed for mode A/C are included in the list, this means that the A/C Mode
item can be built. It is also important to note, when checking the STATNotes list for an item, the item
may exist under the "APOC" name. For example, if you are searching the list for the Set Rate item, it may
be listed as APOC set rate.
If any of the STATNotes items do not exist, create them.
When creating a Mode item, the following steps will be completed:
Item Name
Item Prompt on Analyzer
Description
For this example, do not specify a Description.
Entry Prompt
i-STAT Alinity — System Operations Manual
STATNotes Items
Set Rate
Vt
FIO2
PEEP
PS
Required. Specify a name that is from 1 to 20 characters long and:
• does not begin with the characters APOC (in upper or lower
case)
• is a unique name within this healthcare organization.
Required. Specify a prompt that is from 1 to 20 characters long and:
• does not begin with the characters APOC (in upper or lower
case)
• is a unique name within this healthcare organization.
Optional.
Select one of the following:
Entry, optional
Prompt the operator for an optional entry.
Entry, mandatory with override
Prompt the operator for a mandatory entry. If the override feature
is chosen by the operator no further action is required and the
instrument will advance to the next screen.
Art: 746300-01 Rev. N