Sample Types

Sample Types
Use Default Sample Types
Custom Sample Types
Sample Type Entry Options
When finished, choose from the options shown at the bottom of the screen:
• Previous to go back to the previous screen
• Discard to undo changes
• Finish Later to create a draft
• Next to open the next tab
i-STAT Alinity — System Operations Manual
Unselect this check box to use Custom Sample Types. Default
sample types are:
Arterial
Venous
Capillary
Unspecified
Define a name for each custom sample type. Specify a name up to
20 characters long. A maximum of 6 sample types can be defined. In
the drop down box next to the name, select the sample type: other,
arterial, venous, or capillary. The default sample type that is
selected from the drop down box will provide the graphics help
displays on the instrument while the operator is using the patient
testing pathway.
Determine whether the operator is prompted to enter the type of
sample used.
Enable
Unselect the check box if you do not want to
prompt the operator for the sample type.
Optional
User is prompted to enter optional sample type.
Mandatory with override
User is prompted to enter mandatory sample
type. If the override feature is chosen by the
operator no further action is required and the
instrument will advance to the next screen.
Mandatory
User is required to enter sample type.
Art: 746300-01 Rev. N
181
Rev. Date: 02-Nov-2022