Cartridge Lot QC Settings
Search Cartridge List for Cartridge Lot Number
Enable
Cartridge QC Warning Screen
Enable Cartridge QC Warning
Screen
When finished, choose from the options shown at the bottom of the screen:
• Previous to go back to the previous screen
• Discard to undo changes
• Finish Later to create a draft
• Next to open the next tab
i-STAT Alinity — System Operations Manual
User-defined cartridge QC settings to aid in compliance with
regulatory agency requirements or recommendations. Do not select
the options below if the data management system in use does not
support cartridge lot QC. The data management system must be
able to accept cartridge lot numbers and change the status so that
cartridges are available for use.
Select this check box to search for the lot number in a list of
cartridges in use or ready for use.
Lot Number not on List --
QC Tests
Select this check box to display a warning screen on the instrument
when the i‑STAT cartridge QC is due. If you select this feature, the
following option is available:
Show warning n days in
advance
Art: 746300-01 Rev. N
Determine the action to be taken if
the lot number is not on the list:
Warn
Display a warning to the operator and
allow the test to be run in the QC test
pathway only. Patient and proficiency
testing is disabled until the cartridge
lot number is available in the
cartridge list.
Lockout
Do not allow the test to be run until
the lot number has been added to
the list of cartridges.
Specify the number of days in
advance of the QC deadline when the
warning message will be displayed.
Allowable values range from 1
through 365. The default is 1.