Quality Options: Quality Control
Quality Options: Cal Ver
Quality Options
i-STAT Alinity — System Operations Manual
Perform Electronic
Follow the prompts to perform testing.
Simulator Test
Perform Unscheduled Cal Ver
Scheduled Cal Ver
Perform Proficiency Test
Update eVAS
View Disabled Cartridges
Art: 745527-01 Rev. N
When calibration verification
testing is not scheduled via a
customization profile or if an
additional cal/ver testing event is
desired, this is the testing
pathway. Follow the prompts on
the screen to perform testing
If calibration verification testing
has been set via a customization
profile, this is the testing
pathway. Follow the prompts on
the screen to perform testing.
Follow the prompts on the screen to perform
testing.
Install from USB - Follow prompts on the
screen to install instrument software.
Note: Some preformatted USB flash
drives may not work with the Alinity
system. To avoid issues, reformat the
drive using a Windows PC before
using the USB flash drive with the
Alinity system.
If the instrument does not detect the base
station, try the following:
• remove the instrument and re-seat it on the
Base Station
• check to ensure that the Base Station is
getting power (blue light illuminated)
If the instrument does not detect the USB
drive, try the following:
• remove the USB and reinsert
• ensure that the USB is formatted
Install from Server
Cartridges are disabled due to expired QC.