252
i‑STAT Alinity only displays the last 3 numbers or
letters of the Operator ID or No Operator ID. The
entire Operator ID should display.
i‑STAT Alinity only displays the operator's first
name and last initial. The entire name should
display.
i‑STAT Alinity will not upload operator list
i‑STAT Alinity will not upload cartridge lot list
Patient armband barcode will not scan
Patient list will not upload onto the i‑STAT Alinity
i-STAT Alinity — System Operations Manual
Issue
Art: 746304-01 Rev. H
Solution
Check the Operator ID Presentation feature in the
Operator ID Entry feature on the profile assigned
to the instrument. Follow CWi instructions in this
manual to change the display.
Check the Operator Name Presentation in the
Operator ID Entry feature on the profile assigned
to the instrument. If the check box is marked for
Partially Display Operator Name, uncheck it to
display entire name. Follow CWi instructions in
this manual.
Check and confirm that the Search for Operator ID
on Operator List feature in the Operator Settings
Category is checked on the profile assigned to the
instrument. If not checked, follow CWi instructions
in this manual to change the profile and enable
the instrument to load the operator list from the
data management system. Check the Connectivity
Settings Category to confirm the IP address is
programmed for operator list.
Check to confirm Search Cartridge List for
Cartridge Lot Number is enabled in the Quality
Settings Category on the profile assigned to the
instrument. If not enabled, follow CWi instructions
in this manual to enable Search Cartridge List for
Cartridge Lot Number to load the cartridge list
from the data management system. Check the
Connectivity Settings Category to confirm the IP
address is programmed for cartridge list.
Check Barcode settings in the Patient Settings ID
Entry feature on the profile assigned to the
instrument. Follow CWi instructions in this manual
to change the barcode type.
Check and confirm that the Search Patient List for
Patient ID feature is enabled in the Patient
Settings Category on the profile assigned to the
instrument. If it is not, follow the CWi instructions
in this manual to change the profile and enable
the instrument to load the Patient List from the
data management system. Check the Connectivity
Settings Category to confirm the IP address is
programmed for Patient List.
Rev. Date: 29-Oct-2019
Need help?
Do you have a question about the i-STAT Alinity and is the answer not in the manual?