A. From the Home screen, navigate to: More Options > Instrument Options > Network Settings >
Network Services
B. Enter your Operator ID and sequence through the workflow to view the press Next to view the
Firmware Delivery option. Verify Enabled is selected.
5. The battery charge level is no less than 51%.
To verify:
A. From the Home Screen, navigate to: More Options > Instrument Status
B. Locate Battery and verify level is 51% or higher.
Procedure:
Perform the following steps to download and install software update:
CAUTION: During the installation, do not remove the instrument from the Base Station. When the
installation is complete, the instrument will display a message indicating that new software has been
installed. When installation is complete, the instrument may be removed from the Base Station.
1. From the Home screen, navigate to: More Options > Instrument Options > Software Installation >
Check for Update
2. After the software download is completed, installation of software can begin. Please skip step 3.
3. Software download and installation can be performed at different times if needed. The i‑STAT Alinity
will display the following symbol
software is downloaded but not installed. Touch this icon to display the Software Installation menu.
Touch the Install Pending button to install the software.
4. Follow the on-screen instructions to continue with software installation.
5. When presented with the name of the software package to be installed, record the name before
proceeding (this information may be used at the end of the procedure to verify that the software
update has been completed).
• Touch Next to continue with the installation of the software package, OR
• Touch Cancel to terminate the installation
6. Continue to follow the on-screen instructions. During the installation process, the instrument screen
will go blank and remain blank for several minutes. This is normal.
• Do not remove the instrument from the Base Station
7. When the installation is complete, the instrument will display a message indicating that new software
has been installed. The display of this message is the indication that it is safe to resume normal
operation of the instrument.