Abbott i-STAT Alinity Operation Manual page 121

Hide thumbs Also See for i-STAT Alinity:
Table of Contents

Advertisement

Analyte
iCa (mmol/L)
Lac (mmol/L)
Crea (µmol/L)
PCO
(mmHg)
2
PO
(mmHg)
2
H+ (pH)
Reactive ingredients for i-STAT TriControls Calibration Verification materials
Analyte
Na (mmol/L)
K (mmol/L)
Cl (mmol/L)
Glu (mg/dL)
Urea (mg/dL)
iCa (mmol/L)
Lac (mmol/L)
Crea (mg/dL)
PCO
(mmHg)
2
PO
(mmHg)
2
H+ (pH)
Evaluating Cal Ver Results
Ranges
Calibration throughout the reportable range of each analyte is verified if each analyte value falls within
the corresponding range in the Value Assignment Sheet.
If a result for a level falls outside the assigned range, follow laboratory policy. Contact Technical Support
for troubleshooting information.
Note: If the calibration verification set is to be used to assess linearity, plot the analyte
value against the mean value of the acceptable range. The concentrations of analytes in
the calibration verification set are not intended or prepared to be equally spaced.
The following is an example of the mean and range for level 2 Cal Ver solution taken from a Value
Assignment Sheet:
i-STAT Alinity — System Operations Manual
Cal Ver Level 1
Cal Ver Level 2
2.5
1.6
19.5
8.4
1486
386
95
66
43
61
6.81
7.15
Cal Ver Level 1
Cal Ver Level 2
97
118
2.30
3.00
67
76
595
285
114
44
0.40
0.90
17.7
8.30
15.6
4.65
96
65
40
63
6.550
7.025
Mean
Cal Ver Level 3
Cal Ver Level 4
1.3
0.8
2.3
1
155
46
30
22
100
140
7.41
7.60
Cal Ver Level 3
Cal Ver Level 4
124
150
4.00
6.30
94
119
160
65
8.4
4.6
1.35
1.58
3.00
1.63
1.59
0.65
40
26
120
163
7.390
7.610
Range
Art: 745537-01 Rev. F
83
Cal Ver Level 5
0.2
0.6
17
18
400
7.95
Cal Ver Level 5
159
8.20
134
53
3.0
2.40
1.52
0.55
12
500
7.850
Rev. Date: 02-Nov-2022