Precautions and Limitations

For safety when using the Base Station:
DO:
• use only the accessories and consumables specified or supplied for this system by Abbott Point of
Care.
• use only the power supply shipped in the box with the Base Station. Attempting to use a different
type of manufacturer's adapter could damage the unit and cause fire or explosion hazards.
• be sure to install all cables and power supplies so they do not pose a trip hazard. Mount equipment
so cables and accessories stay clear of walkways.
• always use the Base Station for charging. Refer to the rechargeable battery Getting Started Guide on
instructions for proper charging.
• connect only Abbott Point of Care provided printers to the Base Station printer port.
• check with authorities for local, state, and/or national requirements for disposal.
DO NOT:
• try to connect any non-electrically isolated equipment to the Base Station.
• connect the Base Station to unauthorized medical devices or other equipment.
• place metal objects on or near the exposed battery charging contacts.
• place the instrument in an oxygen-enriched atmosphere.
• make any unauthorized repairs or modifications to this product, as it may cause personal injury or
damage to the unit.
NOTE:
• The AC adapter power supply plug acts as a disconnect device for the Base Station and therefore, the
AC socket outlet must be installed (or located) near the Base Station and must be easily accessible if
the Base Station needs to be unplugged. When correctly connected to power, the blue light on the
Base Station will be illuminated. After an instrument is docked, the blue light above the screen will be
illuminated. This may take several seconds. If the blue light does not illuminate, check the power
supply. Ensure that the Base Station is connected properly and that the instrument is seated correctly
in the Base Station.
• Protection provided by this equipment may be impaired if used in a manner not specified by Abbott
Point of Care.
• The instrument and its peripherals are not listed by any authority with respect to suitability for use in
oxygen-enriched atmospheres.
• The Base Station must be on a flat, level, horizontal surface such as a table top. Falling equipment
may cause injury. Always place the equipment on a stable level surface in a location where it will not
cause injury if dropped.
i-STAT Alinity — System Operations Manual
Art: 745528-01 Rev. G
53
Rev. Date: 16-May-2022