Operator Actions - Abbott i-STAT Alinity Operation Manual

Hide thumbs Also See for i-STAT Alinity:
Table of Contents

Advertisement

154

Operator Actions

The features in this section determine the actions for Operators when working in the patient test
pathway or training mode pathway.
Operator Actions
Accept or Reject Results
Rejection Comment
Rejection Comment Style
i-STAT Alinity — System Operations Manual
Determines whether the operator has the ability to accept or reject
results after they are displayed on the instrument.
Enable Accept or Reject
Results
Select either option:
Optional
Mandatory
Controls display of comments on the i‑STAT Alinity. Select either:
Comment List
Comment List with text
box
Art: 746300-01 Rev. N
Select the check box to allow operators
to accept or reject test results after
results display on the instrument. If
results are rejected no patient
treatment should be performed from
the rejected results. Unselect the check
box and operators will not have the
option of rejecting or accepting results
after they display on the instrument. If
results are transmitted to the data
management system, only those that
are accepted will go through to the LIS/
EMR. Rejected results will not be sent to
the LIS/EMR.
The operator is prompted to enter a
comment for rejecting test results.
The operator is prompted to enter a
mandatory comment for rejecting test
results.
Only comments chosen from the
selection list will be displayed and
available for attaching to a rejected
result.
Comments chosen from the comment
list along with a text box that will accept
20 characters will be displayed and
available for attaching to a rejected
result. A maximum of 6 comments can
be displayed with a text box.
Rev. Date: 02-Nov-2022