Resolutions
N
The cartridge was not fully inserted. Repeat testing with a freshly filled cartridge. Ensure the
cartridge is fully inserted. The cartridge is fully inserted when the click is heard. If the same
quality check failure displays, contact the system administrator for further instruction.
P
The instrument did not reset correctly. Repeat electronic simulator testing. If the test results in a
PASS, the instrument is ready for use otherwise contact the system administrator for further
instruction.
R
Test has been successfully cancelled.
S
OSi software installation required. Contact the system administrator for further instructions
T
Install the most recent OSi software. Contact the system administrator for further instructions.
U
Power off the instrument. Insert the Latch Return Tool into the cartridge port until it stops.
Immediately remove the tool from the instrument. Repeat testing with a freshly filled cartridge.
If the same quality check failure displays, contact the system administrator for further
instruction.
V
Power off the instrument. Insert the Latch Return Tool into the cartridge port until it stops.
Immediately remove the tool from the instrument. Prepare a new bottle of material per the
manufacturer's instruction. Repeat testing with a freshly filled cartridge. If the same quality
check failure displays, contact the system administrator for further instruction.
W
When filling a cartridge, use care to advance blood to the level indicated by the 'fill to' arrow.
Draw new sample. Repeat testing with a freshly filled cartridge. Carefully observe the help
provided throughout the testing pathway. If the same quality check failure displays, contact the
system administrator for further instruction.
X
Draw new sample. Repeat testing with a freshly filled cartridge. Carefully observe the help
provided throughout the testing pathway. If the same quality check failure displays, contact the
system administrator for further instruction.
Y
Cartridge pouch must be out of the refrigerator for a minimum of 5 minutes before opening.
After opening the pouch, immediately begin to follow the prompts on the screen. If the same
quality check failure displays, contact the system administrator for further instruction.
Z
The Conditioning Cartridge was not fully inserted. Repeat conditioning. Ensure the Conditioning
Cartridge is fully inserted. The Conditioning Cartridge is fully inserted when the click is heard. If
the same quality check failure displays, contact the system administrator for further instruction.
AA
Repeat conditioning with Conditioning Cartridge testing. If conditioning completes successfully,
the instrument is ready for use, otherwise contact the system administrator for further
instruction.
BB
Instrument did not reset correctly. Repeat conditioning with Conditioning Cartridge. If
conditioning completes successfully, the instrument is ready for use, otherwise contact the
system administrator for further instruction.
CC
A cartridge was detected when a Conditioning Cartridge was expected. Repeat conditioning
ensuring to insert a Conditioning Cartridge. Ensure the Conditioning Cartridge is fully inserted.
The Conditioning Cartridge is fully inserted when the click is heard. If the same quality check
failure displays, contact the system administrator for further instruction.
i-STAT Alinity — System Operations Manual
Art: 745527-01 Rev. N
39