Abbott i-STAT Alinity Operation Manual page 114

Hide thumbs Also See for i-STAT Alinity:
Table of Contents

Advertisement

76
Analyte
Lac (mmol/L)
Crea (µmol/L)
PCO
(mmHg)
2
PO
(mmHg)
2
H+ (pH)
Reactive Ingredients for i-STAT TriControls Control Solutions:
Analyte
Na (mmol/L)
K (mmol/L)
Cl (mmol/L)
Glu (mg/dL)
Urea (mg/dL)
iCa (mmol/L)
Lac (mmol/L)
Crea (mg/dL)
PCO
(mmHg)
2
PO
(mmHg)
2
H+ (pH)
Evaluating QC Results
Ranges
The following is an example of the mean and range for level 1 control taken from a Value Assignment
Sheet:
Na
Na is the analyte tested, Na (Sodium).
mmol/L, mEQ/L are the units of measurement.
Mean is the mean or average level of sodium (Na).
Range is the acceptable range of sodium (Na) levels.
i-STAT Alinity — System Operations Manual
Control Level 1
8.4
386
66
61
7.15
Control Level 1
118
3.00
76
285
44
0.90
8.30
4.65
65
63
7.025
mmol/L, mEq/L
Control Level 2
2.3
155
30
100
7.41
Control Level 2
124
4.00
94
160
8.4
1.35
3.00
1.59
40
120
7.390
Mean
Range
120
116-125
Art: 745533-01 Rev. H
Control Level 3
1
46
22
140
7.60
Control Level 3
150
6.30
119
65
4.6
1.58
1.63
0.65
26
163
7.610
Rev. Date: 02-Nov-2022