8. Follow the instructions on screen to complete the Electronic Simulator testing, or press Exit Alerts to
resume normal operation of the instrument.
Note: It is recommended that Electronic Simulator Test be completed immediately following a
software update. Depending on the instrument customization, testing may be disabled until an
Electronic Simulator Test with passing result is completed.
Troubleshooting:
If there is any question as to whether or not the software update completed, check the software status
page. From the Home screen, navigate to: More Options > Instrument Options > Software Installation >
Software Status
• If the Firmware name displayed on the Software Status page matches that which was recorded during
the software update procedure, this indicates that the software update was completed.
Symptom
Battery status indicator does not indicate battery
is charging
i-STAT Alinity — System Operations Manual
Action to take
• Remove the instrument and re-seat it on the
Base Station
• Ensure the Base Station is getting power (Blue
light on Base Station is illuminated)
Art: 754496-01 Rev. E
103
Rev. Date: 02-Nov-2022