108
Symptom
Instrument indicates a connection could not be
made to the Server service, OR
Instrument powers off before completing the
software download, OR
Instrument does not power on within 10 minutes
after screen going blank
Repeated attempts to update software via wireless
network connection fail
i-STAT Alinity — System Operations Manual
Action to take
Verify the following as described in the
prerequisites, and repeat the software update
procedure:
• The wireless signal strength is 3 solid bars or
greater
• Instrument is reporting a valid IP address for
the wireless connection
• i-STAT Alinity instrument(s) with appropriate
Network Configuration file installed
• The healthcare organization network is
allowing the instrument to connect to the
internet in order to have access to the Abbott
Managed Server or to connect to the SDi
Follow the procedure to perform software update
using USB memory device
Art: 754496-01 Rev. E
Rev. Date: 02-Nov-2022