Abbott i-STAT Alinity Operation Manual page 191

Hide thumbs Also See for i-STAT Alinity:
Table of Contents

Advertisement

Decimal Character
Memory Full Action
The instrument can store 500 results (including Patient, QC, Training results). This setting controls the
action taken when the memory is full:
When finished, choose from the options shown at the bottom of the screen:
• Previous to go back to the previous screen
• Discard to undo changes
• Finish Later to create a draft
• Next to open the next tab
i-STAT Alinity — System Operations Manual
Select the decimal point or comma for the display of results on the
instrument. Decimal point is the default.
Overwrite the oldest record
without notification
Memory full notification at
instrument start-up
Lockout all Testing Pathways until
memory is cleared
Art: 746300-01 Rev. N
New records overwrite old
records. The oldest record is
overwritten first, the next oldest
record is overwritten next, and
so on. This is the default.
When the instrument is powered
up a start up alert will display
indicating the memory is full.
The operator is prompted to
clear the memory, otherwise old
records may be overwritten.
Testing will be suspended until
the memory is cleared. Lockout
message displays on the
instrument at start up.
Rev. Date: 02-Nov-2022