Abbott i-STAT Alinity Operation Manual page 293

Hide thumbs Also See for i-STAT Alinity:
Table of Contents

Advertisement

Analyte ranges were created with age and gender
applied but no reference ranges or action ranges
will display on the i‑STAT Alinity. The display shows
no color.
STATNotes set does not display when clicking on
the drop down arrow beside the cartridge.
Category created does not appear in the list of
available categories when creating a profile
Cartridge QC Lot List is applied and defined in CWi
and data management system but QC or Cal Ver
results are not being displayed in the data
management system.
i-STAT Alinity — System Operations Manual
Issue
Art: 746304-01 Rev. H
Solution
Check to confirm in Patient Settings, Basic
Functions that Patient Age/Gender Entry has
been enabled. If not, follow the directions in this
manual to change setting and upload profile with
correction.
A STATNotes set must be created and published
under the Manage Set tab before the set will
appear as available for application to a cartridge.
Follow the instructions in this manual to create a
STATNotes set.
Categories must be in the Published status to
appear as available when creating a profile.
CWi Pass/Fail determination for QC or Cal Ver is
required. If set to None the data management
system will not recognize the test performed.
Follow the instructions in this manual to set Auto
via eVAS or Manual.
253
Rev. Date: 29-Oct-2019

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents