i‑STAT cartridges contain test reagents which are located on sensors on the top of the cartridge. The
instrument and cartridge work together to generate a clinically meaningful result.
The cartridges are designed for use with the i‑STAT Alinity instrument.
Note:
• Not all cartridges are available in all regions. Check with your local representative for
• Customization may affect analyte availability.
• For access to cartridge-specific Instructions for Use (IFU) and analyte-specific CTI
Cartridge Specifications
Shelf Life: Refrigerated at 2ºC to 8ºC (35ºF to 46ºF) until expiration date. Refer to the cartridge box for
room temperature storage requirements.
Preparation for Use: Individual cartridges may be used after standing five minutes at room temperature.
An entire box of cartridges should stand at room temperature for one hour. All cartridges should be used
immediately after opening pouch. If the pouch has been punctured, the cartridge should not be used.
Cartridges stored in refrigerator:
• Temperature must be 2-8°C (35-46°F)
• Cartridges expire on the date printed on the pouch
Cartridges stored at room temperature:
• Temperature must be 18-30°C (64-86°F)
• Once cartridge is at room temperature the expiration date changes
• Cartridge should not be returned to the refrigerator once out for more than 5 minutes.
i-STAT Alinity — System Operations Manual
availability in specific markets. Scanning a cartridge that is unavailable will result in a
pop-up window that displays Invalid Cartridge Type.
Sheets, access the Support page on the Abbott Point of Care website
www.globalpointofcare.abbott.

1.12 - i-STAT Cartridges

Art: 745524-01 Rev. I
85
Rev. Date: 02-Nov-2022