8. Follow the instructions on screen to complete the Electronic Simulator testing, or press Exit Alerts to
resume normal operation of the instrument.
Note: It is recommended that Electronic Simulator Test be completed immediately following a
software update. Depending on the instrument customization, testing may be disabled until an
Electronic Simulator Test with passing result is completed.
Troubleshooting:
If there is any question as to whether or not the software update completed, check the software status
page. From the Home screen, navigate to:More Options > Instrument Options > Software Installation >
Software Status
• If the Firmware name displayed on the Software Status page matches that which was recorded during
the software update procedure, this indicates that the software update was completed.
Symptom
Check for Update button is not enabled
i-STAT Alinity — System Operations Manual
Action to take
Verify the following as described in the
prerequisites, and repeat the software update
procedure:
• The instrument is running software version
OSi05 or later
• Network Services for software update enabled
on instrument
• The battery charge level is no less than 51%
Art: 754496-01 Rev. E
107
Rev. Date: 02-Nov-2022