Interpreting Results - Abbott i-STAT Alinity Operation Manual

Hide thumbs Also See for i-STAT Alinity:
Table of Contents

Advertisement

On the right side of the screen are action buttons for:
touch to display the Home screen.

Interpreting Results

The following section provides examples for various analytes. Not all cartridges are available in all
regions. Ability to run a cartridge is controlled by the instrument and by region.
Calculated Result Indicator
A star (*) beside an analyte indicates a calculated result, as shown in the example below:
Suppressed Results - Stars (***)
Various conditions cause results to be suppressed. The instrument displays stars (***), referred to as a
"star out", as shown in the example below:
Reasons such as those listed below may produce starred out results:
• an uncharacteristic sensor signal
• a bad sensor
• improperly stored cartridge
i-STAT Alinity — System Operations Manual
When customized, Transmits result to
Data Manager
On the bottom of the screen are the icons:
Art: 745531-01 Rev. G
touch to display the next screen.
Rev. Date: 04-Mar-2021