Electronic Simulator

Testing for the external Electronic Simulator can be scheduled to be performed based on an interval of a
number of days. This setting is usually determined by compliance with regulatory agencies.
Simulator Testing Schedule
Run Electronic Simulator every n
days
When finished, choose from the options shown at the bottom of the screen:
• Previous to go back to the previous screen
• Discard to undo changes
• Finish Later to create a draft
• Next to open the next tab
i-STAT Alinity — System Operations Manual
Note: Internal Electronic Simulator testing is performed every time a cartridge is tested
on the i‑STAT Alinity.
Determines how frequently the external electronic simulator is
tested. Specify a number of days in the range 0 - 365. The default is
0, which turns off testing.
Note: Abbott Point of Care manufacturer's requirement is
for the electronic simulator to be tested every 6 months
(182 days).
Art: 746300-01 Rev. N
Rev. Date: 02-Nov-2022