ACT (Activated Clotting Time)

This screen allows you to customize the setting for ACT testing with Kaolin activator. You can select
between the current 37° (Pre-warm) result calibration and an ambient temperature (Non-warm) result
calibration.
Options on the screen are:
ACTk Mode
Stop ACT Test
When finished, choose from the options shown at the bottom of the screen:
• Previous to go back to the previous screen
• Discard to undo changes
• Finish Later to create a draft
• Next to open the next tab
i-STAT Alinity — System Operations Manual
For ACT-K (Kaolin ACT) cartridges, select either:
Pre-warm
37°C result calibration
applies to patient test pathway only
pre-warm is the default
Non-warm
ambient temperature result calibration
Determines whether an operator can stop an ACT test while it
is being performed. Select the check box to enable stopping a
test. This is the default. Unselect the check box to prevent an
operator from stopping a test.
Art: 746300-01 Rev. N
189
Rev. Date: 02-Nov-2022