Specifications

Dimensions: Height x Width x Depth
Weight
Electrical rating
Operational range: Temperature and humidity 10 to 40ºC (50 to 104ºF)
Storage range: Temperature and humidity
Altitude

Precautions and Limitations

Batteries have been designed to provide a safe, high-capacity power source in a relatively small
lightweight package. However, if misused or abused, these batteries can be dangerous. Follow these
guidelines for the safe handling, use and disposal of the rechargeable batteries.
DO:
• refer to the Getting Started Guide for proper charging instructions.
• charge a new rechargeable battery pack for 4 hours before initial use. A fully discharged battery will
be 100% charged and ready for use after 4 hours.
• use only a rechargeable battery pack purchased from Abbott Point of Care.
• use only the accessories and consumables specified or supplied for this system by Abbott Point of
Care.
• keep a spare, charged battery on hand at all times.
• store charged battery in the original packaging.
• check with authorities for local, state, and/or national requirements for disposal or recycling
requirements for Lithium-Ion batteries.
DO NOT:
• dismantle, open or shred the battery.
• expose the battery to heat or fire. Avoid storage in direct sunlight.
• short circuit the battery. Do not store batteries in such a manner that they may short circuit each
other, or allow metal objects to touch the battery contacts.
• subject the batteries to mechanical shock.
• open, disassemble, modify, crush, burn, shred, or expose to high temperatures. Toxic gases and
liquids are present in batteries.
• place the battery in an oxygen-enriched atmosphere.
• make any unauthorized repairs or modifications to this product as this may cause personal injury or
damage to the unit.
i-STAT Alinity — System Operations Manual
i-STAT Alinity Rechargeable Battery Specifications
2.7 in (6.9 cm) x 3.6 in (9.1 cm) x 1.9 in (4.8 cm)
0.4 lb (180 g)
3.65 VDC 19.3 Wh (nominal)
10 to 90% non-condensing relative humidity, with
maximum saturation temperature of 34ºC (93.2ºF)
-10 to 60ºC (14 to 140ºF)
10 to 90% non-condensing, with maximum saturation
temperature of 50ºC (122ºF)
up to 3,048 meters (10,000 feet)
Art: 745529-01 Rev. G