Download Print this page
Abbott CentriMag Quick Reference Manual

Abbott CentriMag Quick Reference Manual

Circulatory support system alarms and alerts
Hide thumbs Also See for CentriMag:

Advertisement

Quick Links

CentriMag
Circulatory Support System
TM
Alarms and Alerts Quick Reference Guide
Alarm/
Text Mes-
ID
System Status & OperatorResponse
Alert
sage
The Pump will not start. An audible alarm will sound, which
cannot be muted. Switch the Console OFF and ON again. If
POWER ON
S1
Alarm
the alarm reappears, switch to the backup Console, Motor
TEST FAIL
and Flow Probe, record the alarm message and contact
your local Abbott Medical representative.
The Pump will stop. An audible alarm will sound, which
SYSTEM
cannot be muted. Clamp the return tubing and switch to
FAULT
the backup Console, Motor and Flow Probe according the
S2
Alarm
(Run-Time
procedure described in Section 2. Resume support. Record
System
the alarm message and contact your local Abbott Medical
Failure)
representative.
The Pump will stop. An audible alarm will sound, which
can be muted for 60 seconds.Clamp the return tubing
MOTOR
and switch to the backup Console, Motor and Flow Probe
M1
Alarm
STOPPED
according the procedure described in Section 2. Resume
support. Record the alarm message and contact your local
Abbott Medical representative.
The Pump will stop. An audible alarm will sound, which can
be muted for 60 seconds.
During setup of the System:
Press the alarm acknowledge button and check that the Mo-
tor connector is fully inserted into the back of the Console.
During support:
MOTOR
Press the alarm acknowledge button and check that the
M2
Alarm
DISCON-
connector of the Motor is fully inserted into the back of the
NECTED
Console. Resume support. If the visual alarm message
does not disappear, clamp the return tubing and switch to
the backup Console, Motor and Flow Probe according the
procedure described in Section 2. Resume support. Record
the alarm message and contact your local Abbott Medical
representative.
System will not start. An audible alarm will sound, which can
be muted for 60 seconds. Press the alarm acknowledge
PUMP NOT
button. Insert or reinsert the pump and secure it with
M3
Alarm
INSERTED
the locking screw. If the alarm repeats, switch to backup
Console, Motor and Flow Probe according the procedure
described in Section 2.
An audible alarm will sound and the System will continue
to operate. Press the alarm acknowledge button, if the
visual alarm message does not disappear, clamp the return
MOTOR
M4
Alarm
tubing, stop the pump and switch to the backup Console,
ALARM
Motor and Flow Probe according the procedure described
in Section 2. Resume support. Record the alarm message
and contact your local Abbott Medical representative.
Check the Pump flow. If pump flow is satisfactory, reduce
set speed while insuring that flow is maintained. Press
the alarm acknowledge button. If the alert repeats, clamp
the return tubing, stop the pump and switch to the backup
Console, Motor and Flow Probe according the procedure
SET PUMP
described in Section 2. Resume support. Record the alert
M5
Alert
SPEED NOT
message and contact your Abbott Medical representative.
REACHED
If the pump flow is not satisfactory, clamp the return tubing
and switch to the backup Console, Motor and Flow Probe
according to the procedure described in Section 2. Resume
support. Record the alert message and contact Abbott Med-
ical representative.
1
RESPONSE TO SYSTEM ALARMS OR ALERTS
When a System alarm or alert condition occurs, an alarm or alert tone sounds and
text message appears on the display of the 2nd Generation CentriMag Primary
Console. Most alarms and alerts require some action on the user's part to correct
the cause. The following table can be used to determine how to respond to an
alarm or alert condition.
Alarm/
Text Mes-
ID
System Status & OperatorResponse
Alert
sage
The Console battery will not function. An audible alarm
BATTERY
will sound. Switch to the backup Console, Motor and Flow
B1
Alert
MODULE
Probe according to the procedure described in Section 2.
FAIL
Resume support. Record the alert message and contact
your Abbott Medical representative.
The Pump will stop after a very short time. Plug the Console
BATTERY
into a AC power outlet to charge the battery. If no AC outlet
B2
Alert
BELOW
is available, switch to the backup Console, Motor and Flow
MINIMUM
Probe according to the procedure described in Section 2.
Resume support.
Check the Flow Probe connection on back of the Console.
FLOW
If necessary, reconnect the Flow Probe connector to the
PROBE
F1
Alert
back of the Console. Press the alarm acknowledge button.
DISCON-
Switch to the backup Flow Probe, if the alert message
NECTED
repeats.
Press the alarm acknowledge button, if the message does
not disappear, clamp the return tubing, stop the pump
SYSTEM
and switch to the backup Console, Motor and Flow Probe
S3
Alert
ALERT
according the procedure described in Section 2. Resume
support. Record the alert message and contact your local
Abbott Medical representative.
Manually disconnect, reposition and reconnect the
FLOW
Flow Probe transducer to the tubing. Press the alarm
SIGNAL
acknowledge button. Switch to the backup Flow Probe,
INTER-
if the alert message repeats. If problem still persists after
F2
Alert
RUPTED
switching to the backup Flow Probe, stop the pump and
(Flow rate
switch to a backup Console. Follow the instructions
sensor error)
described in Section 2. Resume support. Record the alert
message and contact your Abbott Medical representative.
FLOW
Check for physiologic cause or circuit obstruction.
BELOW
Check minimum flow set point. Do not increase RPM
F3
Alert
MINIMUM
without confirming that adequate blood volume is available.
(Low Flow)
Common cause of this alert is inadequate blood volume at
the drainage cannula site for the desired pump flow.
FLOW
F4
Alert
ABOVE
Reduce Pump speed and check for cause.
MAXIMUM
Check the electrical connections on the pressure 1 trans-
PRESSURE
ducer and recalibrate. If the problem persists disconnect,
P1
Alert
1 DISCON-
reconnect, and recalibrate the transducer. Consider chang-
NECTED
ing the transducer and cable if the problem persists.
Check the electrical connections on the pressure 2 trans-
PRESSURE
ducer and recalibrate. If the problem persists disconnect,
P2
Alert
2 DISCON-
reconnect, and recalibrate the transducer. Consider chang-
NECTED
ing the transducer and cable if the problem persists.
Alarm/
Text Mes-
ID
System Status & OperatorResponse
Alert
sage
The pressure monitoring system will not function. If pressure
monitoring is needed then change to the backup Console,
PRESSURE
Motor and Flow Probe. Switch to the backup System
P3
Alert
SYSTEM
according the procedure described in Section 2. Resume
FAIL
support. Record the alert message and contact your Abbott
Medical representative.
Check for and resolve a physiological or mechanical cause.
PRESSURE
Ensure appropriate pressure alarms are set. Consider re-
P4
Alert
1 BELOW
calibrating the transducer, if alert cannot be explained by
MINIMUM
conventional troubleshooting.
Check for and resolve a physiological or mechanical cause.
PRESSURE
Ensure appropriate pressure alarms are set. Consider re-
P5
Alert
2 BELOW
calibrating the transducer, if alert cannot be explained by
MINIMUM
conventional troubleshooting.
Check for and resolve a physiological or mechanical cause.
PRESSURE
Ensure appropriate pressure alarms are set. Consider re-
P6
Alert
1 ABOVE
calibrating the transducer, if alert cannot be explained by
MAXIMUM
conventional troubleshooting. Consider reducing RPM to
reduce the pressure if appropriate.
Check for and resolve a physiological or mechanical cause.
PRESSURE
Ensure appropriate pressure alarms are set. Consider re-
P7
Alert
2 ABOVE
calibrating the transducer, if alert cannot be explained by
MAXIMUM
conventional troubleshooting. Consider reducing RPM to
reduce the pressure if appropriate.
Switch to backup Console, Motor and Flow Probe accord-
MOTOR
M6
Alert
ing to the procedure described in Section 2. Verify that the
OVER TEMP
backup Motor stands free and is not covered (e.g. blankets).
Press the alarm acknowledge button. If the alert message
BATTERY
repeats, switch to backup System as described in Section
B3
Alert
CHARGER
2. If this alert is associated with BATTERY MODULE FAIL
FAIL
then carry out the procedure associated with that alarm.
BATTERY
Do not use the Console. Perform battery maintenance ac-
MAINTE-
B4
Alert
cording to the Battery Maintenance Procedure listed in the
NANCE
CentriMag Circulatory Support System Operation Manual.
REQUIRED
Plug the Console into a AC power outlet to charge the
battery. If no AC outlet is available, switch to the backup
LOW
B5
Alert
Console, Motor and Flow Probe according to the procedure
BATTERY
described in Section 2.
Resume support.
Verify that the user wants the Console to be on battery. If so,
ON BAT-
carefully monitor the status of the battery charge indicator,
B6
Alert
TERY
while using the System on battery.
Reconnect to AC-outlet, as soon as possible.
PL-0370 Rev 00 / December 2019 / Page 1

Advertisement

loading

Summary of Contents for Abbott CentriMag

  • Page 1 When a System alarm or alert condition occurs, an alarm or alert tone sounds and Alarms and Alerts Quick Reference Guide text message appears on the display of the 2nd Generation CentriMag Primary Console. Most alarms and alerts require some action on the user’s part to correct the cause.
  • Page 2 If the retaining screw is not visible in a notch, loosen the retaining screw, remove the pump, remount, rotate the Pump For additional information, refer to the CentriMag Circulatory Support System counterclockwise, and secure by advancing the retaining screw.