c. Locate the two red fittings supplied with the cell. Use one fitting
d. Install a double-cone 10-32 ferrule fitting (P/N 043276) and 10-
4. Connect a backpressure line to the cell outlet (recommended to
prevent bubbles from forming in the cell):
a. Cut a length of 0.25-mm (0.010-in) ID black PEEK tubing
b. Use the red fitting supplied with the cell to connect one end of the
5. Connect a waste line to the backpressure line:
a. Locate the 0.125-mm (0.020-in) ID orange PEEK tubing in the
b. Install a 10-32 ferrule fitting and bolt on one end of the tubing.
c. Use a union (P/N 042627) to connect this line to the backpressure
6. Push the inlet and outlet line unions onto the clips.
7. Route the lines to the outside of the detector through the slot provided
in the enclosure (see
container.
Notes for Cutting Tubing
Use a tubing cutter to cut tubing to the required length. Make sure the
cut is at a right angle to the length of the tubing and there are no nicks
or burrs on the end.
Tighten all fittings fingertight. After starting operation, check for
leaks and tighten more only if a fitting is leaking. If you use a wrench,
Doc. 065377-01 1/12
peak broadening effects due to excessive dead
volume.
to connect the inlet tubing to the inlet union. Note: These fittings
do not use ferrules.
32 bolt (P/N 043275) on the other end of the inlet tubing and
connect it to the separator column outlet.
(P/N 052306) to the required length: 1 meter for a flow rate of
1.0 mL/min, 2 meters for a flow rate of 0.5 mL/min, and so on.
tubing to the outlet union.
Dionex VWD Ship Kit (P/N 064787). Cut a length long enough
to connect the backpressure line to the waste container.
line.
Figure
2-3). Direct the waste line to a waste
B • Installation
87