2. Verify that the detector is correctly configured:
In the Chromeleon 7 Instrument Configuration program or
Chromeleon 6.8 Server Configuration program, go to the General tab
page of the Dionex VWD Properties dialog box.
Turn off Virtual Mode, if it is on.
Click the Browse button and select the Dionex VWD from the Device
List.
The Warning threshold must be higher than the Limit
threshold.
This message appears if you attempt to set a Warning value for certain
parameters (for example, UVLampRelIntensity or VISLampRelIntensity)
that is lower than the Limit value. Chromeleon issues a warning message in
the Audit Trail when the Warning value is reached and an error message when
the Limit value is reached.
To troubleshoot:
Select a Warning value that is higher than the Limit value.
The Warning threshold must be lower than the Limit
threshold.
This message appears if you attempt to set a Warning value for certain
parameters (for example, UVLampOperationTime or
VISLampOperationTime) that is higher than the Limit value. Chromeleon
issues a warning message in the Audit Trail when the Warning value is
reached, and an error message when the Limit value is reached.
To troubleshoot:
Select a Warning value that is lower than the Limit value.