To connect the Dionex VWD to an internal hub on another module:
NOTE If you are installing a Dionex VWD with a Dionex
The USB standard limits the USB cable length to 5 meters
(5.5 yards). Each USB device can be separated from the PC by
no more than five hubs. Thus, if five hubs are installed, each
USB device can be located no more than 30 meters (32 yards)
from the PC.
1. Plug the "A" connector of the USB cable (P/N 063246) into the USB
port on the module with the internal hub.
2. Plug the "B" connector of the USB cable into the USB receptacle on
the Dionex VWD rear panel (see
Dionex VWD Connected to Another Module in the System
Doc. 065377-01 1/12
ICS-5000 Ion Chromatography System, refer to the
system installation instructions for information
about USB compatibility issues with Dionex ICS-
5000 systems.
Module with
VWD
B
USB Cables
Figure B-12. Example USB Connection:
Figure
B-12).
Internal
USB Hub
A
B
PC
A
B • Installation