tighten by small increments (about 1/16th of a turn), just until the leak
stops. Severe overtightening of the bolt can damage the port.
B.5 Connecting the Drain Tube
Connect the drain tube (P/N 055075) to the drain port on the Dionex VWD rear
panel (see
Figure
the level of the workbench on which the Dionex VWD is installed.
The drain tube must remain below the drain port. If any part of the
drain tube is above the drain port, the drip tray may overflow inside
the detector.
88
B-6). Place the free end of the tube in a waste container below
Figure B-6. Dionex VWD Rear Panel
Drain Port
Doc. 065377-01 1/12