Error opening VWD-xxx@USB-xxxx – The System cannot find
the file specified.
Error reading from VWD-xxxx@USB-xxxx.
These errors occur if the USB connection between the detector and the
Chromeleon instrument controller or server is interrupted, or the power
supply to the detector is interrupted.
To troubleshoot:
1. Check the USB connection (see
2. Check the power connection to the detector (see
Error issuing control request to VWDxxx@USB-xxxxx.
The Chromeleon instrument controller or server cannot connect to the
specified detector.
To troubleshoot:
1. Check the USB connection (see
2. Check the power connection to the detector (see
3. Verify that the detector is correctly configured in the Chromeleon 7
instrument or Chromeleon 6.8 timebase:
•
In the Chromeleon 7 Instrument Configuration program or
Chromeleon 6.8 Server Configuration program, go to the General tab
page of the Dionex VWD Properties dialog box.
•
Turn off Virtual Mode, if it is on.
•
Click the Browse button and select the Dionex VWD from the Device
List.