B.4 Installing the Flow Cell
A dummy flow cell is installed in the detector before shipping. Before initial
operation of the detector, replace the dummy cell with a flow cell.
NOTE Keep the dummy flow cell for reuse. The dummy cell
Removing the Dummy Cell
1. Grasp the Dionex VWD front cover by the sides and pull it straight off to
remove it.
2. Loosen the two screws that hold the dummy cell in position (see
and pull the cell out of the detector.
Loosen these
screws
84
protects the detector optics during long-term shutdown
or if the detector must be shipped at a later time.
Figure B-2. Removing the Dummy Cell
Figure
B-2)