Unsuitable flow cell position.
This error may occur when the flow cell is incorrectly installed, thereby
causing the deviation of the calibrated wavelength to be too high for the
measurement and reference channels.
To troubleshoot:
Make sure the flow cell is installed correctly (see
UV lamp ignition failed.
This message may appear if the deuterium lamp fails during operation.
To troubleshoot:
1. Turn the lamp off and on again.
2. Replace the deuterium lamp (P/N 066347) (see
UV lamp intensity low.
This error occurs when the intensity of the deuterium lamp is too low,
compared to the reference intensity.
To troubleshoot:
1. If your application is not overly sensitive to noise, reduce the value of the
lamp relative intensity limit (UVRelIntensity.Limit command) in the
Chromeleon 7 Command window or Chromeleon 6.8 Commands dialog
box.
The default limit is 50%. Some applications can be run when the lamp
intensity is as low as 20%. However, other applications are extremely
sensitive to noise and the limit may need to be set higher than the default.
2. Replace the deuterium lamp (P/N 066347) (see
3. If the error appears again, contact Technical Support for Dionex products
for assistance. The detector optics may be defective.
Doc. 065377-01 1/12
4 • Troubleshooting
Section
5.6).
Section
5.3).
Section
5.3).
51