Dionex VWD Operator's Manual
Figure B-5
To Separator Column
1. Push the flow cell into the opening in the detector and tighten the two
screws.
2. While holding the inlet union in place (see
remove the plug on the end of the union. Repeat to remove the plug
from the outlet union.
3. Connect the flow cell inlet tubing:
a. Locate the required tubing in the Dionex VWD Ship Kit
b. Cut a length of tubing long enough to connect the union on the
86
is a schematic of the flow cell tubing connections.
Cell Inlet Tubing
Cell Outlet
Backpressure Tubing
Cell Waste Tubing
To Waste
Figure B-5. Dionex VWD Flow Schematic
NOTE Be careful not to twist, kink, or otherwise put
stress on the factory-installed lines and
connectors to the flow cell
(P/N 064787). For a standard bore system, locate the 0.25-mm
(0.010-in) ID black PEEK tubing (P/N 052306). For a microbore
system, locate the 0.125-mm (0.005-in) ID red PEEK tubing
(P/N 052311).
flow cell inlet to the separator column outlet.
NOTE Tubing connections between the column and the
flow cell should be as short as possible to avoid
Inlet Union
Outlet Union
Figure
B-5), unscrew and
and
ports.
IN
OUT
Doc. 065377-01 1/12