Dionex VWD Operator's Manual
DAC 2 configured, but not present. Please check
configuration using the Server Configuration program.
This error occurs when a DAC plug-in module is configured in the
Chromeleon 7 instrument or Chromeleon 6.8 timebase, but the software
cannot establish communication with it.
To troubleshoot:
1. Verify that the DAC plug-in module is installed correctly (see
Section
2. If a DAC plug-in module is not installed in the detector, remove it from
the configuration:
In the Chromeleon 7 Instrument Configuration program or Chromeleon
6.8 Server Configuration program, go to the Detector tab page of the
Dionex VWD Properties dialog box. Clear the DAC Board check box
and click OK.
DAC 1 not configured, but present. Please check
configuration using the Instrument Configuration program.
DAC 1 not configured, but present. Please check
configuration using the Server Configuration program.
DAC 2 not configured, but present. Please check
configuration using the Instrument Configuration program.
DAC 2 not configured, but present. Please check
configuration using the Server Configuration program.
This error occurs when a DAC plug-in module is installed in the detector, but
it is not configured in the Chromeleon 7 instrument or Chromeleon 6.8
timebase.
To troubleshoot:
In the Chromeleon 7 Instrument Configuration program or Chromeleon 6.8
Server Configuration program, go to the Detector tab page of the Dionex
VWD Properties dialog box. Select the DAC Board check box and click OK.
46
B.7).
Doc. 065377-01 1/12