Dionex VWD Operator's Manual
To connect the Dionex VWD to an external USB hub:
The Dionex VWD Ship Kit includes one USB cable (P/N 063246). You
must order another USB cable for this configuration.
The USB standard limits the USB cable length to 5 meters
(5.5 yards). Each USB device can be separated from the PC by
no more than five hubs. Thus, if five hubs are installed, each
USB device can be located no more than 30 meters (32 yards)
from the PC.
Carefully secure all USB cables, the USB hub, and the hub
power cable so that they cannot be accidentally disconnected.
1. Plug the "A" connector of a USB cable into a port on the external
USB hub (see
2. Plug the "B" connector of the cable into the USB receptacle on the
Dionex VWD rear panel.
3. Plug the "A" connector of a USB cable into a USB port on the PC.
4. Plug the "B" connector of the cable into a receptacle on the USB hub.
USB Device
Multiple Modules Connected via an External USB Hub
96
Figure
B-13).
VWD
B
USB Cables
B
A
A
A
USB Hub
B
Figure B-13. Example Connections:
USB
Device
B
A
PC
Doc. 065377-01 1/12