Thermo Scientific Dionex ICS Series Operator's Manual page 75

Vwd variable wavelength detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

6. Squeeze the clip on the deuterium lamp connector (see
out the connector from its source.
Deuterium Lamp
Figure 5-2. Deuterium Lamp, Connector, and Clip
7. Loosen the two screws in the lamp flange (see
lamp.
Loosen these
screws
Figure 5-3. Deuterium Lamp Screws and Locating Pin
8. Inspect the new deuterium lamp (P/N 066347) for fingerprints and dust. If
necessary, clean the lamp with isopropanol (IPA).
Doc. 065377-01 1/12
Lamp Connector
5 • Service
Figure
5-2) and pull
Connector Clip
Figure
5-3) and pull out the
Locating Pin
67