For communication between the Dionex Easion and the PC on which Chromeleon is installed, the Dionex Easion must be connected to a USB (Universal Serial Bus) port on either the PC or an external USB hub. For more information, refer to Thermo Scientific Dionex Easion Ion Chromatography System Installation Instructions (Document No.
Dionex Easion Operator’s Manual User Documentation 1.2.1 About This Manual The electronic version (the PDF file) of the Dionex Easion operator’s manual contains numerous hypertext links that can take you to other locations within the file. These links include: •...
People’s Republic of China The Dionex Easion is designed for IC (ion chromatography) applications and should not be used for any other purpose. Operation of a Dionex Easion in a manner not specified by Thermo Fisher Scientific may result in personal injury.
Page 13
1 • Introduction Messages d'avertissement en français Signale une situation de danger immédiat qui, si elle n'est pas évitée, entraînera des blessures graves à mortelles. Signale une situation de danger potentiel qui, si elle n'est pas évitée, pourrait entraîner des blessures graves à mortelles. Signale une situation de danger potentiel qui, si elle n'est pas évitée, pourrait entraîner des blessures mineures à...
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with applicable domestic and international regulations. When the Dionex Easion is delivered to you, it meets all pertinent international electromagnetic compatibility (EMC), safety, and material compliance requirements and directives.
1 • Introduction 1.4.1 Notice on Lifting and Handling of Thermo Scientific Instruments For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/or bulky for one person alone to handle safely.
Page 16
Dionex Easion Operator’s Manual Conformité DEEE Ce produit est conforme avec la directive européenne (2012/19/EU) des Déchets d'Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant: Ce symbole indique que l'équipement ne doit pas être jeté avec les déchets ordinaires, mais doit être collecté...
2 • Description Operating Features 2.1.1 Front Door and Top Cover Figure 2-1 illustrates the front door and top cover of the Dionex Easion. Regenerant Bottle Eluent Bottle Status LEDs Injection Port Figure 2-1. Dionex Easion Doc. 155028-01 08/20...
Page 18
Dionex Easion Operator’s Manual LEDs Three status LEDs (described below) are on the Dionex Easion front door. LED Label If On (Green) If Flashing Power Dionex Easion power is on Does not flash Ready System check passed, but System check failed (occurs if...
Regenerant renews the ability of the suppressor to suppress eluent conductivity. A Dionex Easion configured for anion analyses uses dilute sulfuric acid regenerant. A Dionex Easion configured for cation analyses uses tetrabutylammonium hydroxide (TBAOH) regenerant. For more information about suppressor regeneration, see Section 2.3.5.
Page 20
Section 2.3.2. Pump Heads The Dionex Easion includes a dual-piston serial pump. The flow rate can be set from 0.01 mL/min to 5.00 mL/min. However, for optimum performance, set the flow rate to between 0.20 and 3.00 mL/min. Setting the flow rate to 0.00 mL/min turns off the pump. For more information...
The USB connector connects the Dionex Easion to the PC on which Chromeleon software is installed. For the standard system configuration of one Dionex Easion connected to the PC, connect a USB cable between the USB connector and a USB port on the PC. For more information, refer to the Dionex Easion installation instructions.
Page 22
Dionex Easion and the PC on which Chromeleon is installed. LED Status Description Although the Dionex Easion and the PC are linked, no data is currently being transmitted or received. Flashing The Dionex Easion and the PC are linked and data is being transmitted.
Page 23
2 • Description Plumbing and Waste Lines The following lines exit the Dionex Easion through the tubing chase in the lower-left corner of the rear panel: • Eluent • Regenerant • Cell outlet • Waste (sample, regenerant, and pump priming)
System Component Details 2.3.1 Pump The Dionex Easion pump is a microprocessor-based isocratic eluent delivery system. Its variable speed, dual-piston series design ensures pulse-free pumping for the most demanding applications. Primary Pump Head...
Pressure limits are selected in the Chromeleon Instrument Configuration Manager (in the Dionex Easion Properties dialog box). For troubleshooting guidance if a pressure limit is exceeded, see Section 4.7.
2 • Description Pulse Damper Flow output from the pressure transducer continues to the pulse damper, which smooths minor pressure variations. From there, flow is directed to the injection valve and then to the remainder of the chromatography system. 2.3.3 Injection Valve with Sample Loop The injection valve is a six-port, electrically-activated Rheodyne valve.
Dionex Easion Operator’s Manual column, carrying the contents of the sample loop with it. For a description of how to inject samples manually, see Section 3.11.2. For a description of how to inject samples with an autosampler, see Section 3.11.4.
Temperature compensation further improves baseline stability. Direct conductive heating is used in the Dionex Easion conductivity cell to provide temperature control and compensation. A heat exchanger inside the cell regulates the temperature. All data is collected at 40 °C (104 °F).
Chromeleon Help. Chromeleon Software The Chromeleon 7 Chromatography Data System (version 7.2.10 MUa and later) is used to control the Dionex Easion. Chromeleon provides complete instrument control, data acquisition, and data management. 2.4.1 The Chromeleon ePanel Set The Chromeleon ePanel Set provides access to detailed status and control functions for each system component (pump, detector, and so on).
Figure 2-10 shows ePanel features related to the Dionex Easion. Figure 2-10. Dionex Easion: Chromeleon ePanel Status and Control Functions 2.4.2 Software Control Modes Two modes of software control are available: direct control and programmed control.
2 • Description 2.4.3 System Wellness System Wellness monitors the overall “health” of a chromatographic system. It provides built-in diagnostic and calibration features that help prevent unscheduled system shutdowns and assure reliable operation of system devices. For more information, see Section 5.1.
3 • Operation and Maintenance This chapter describes routine operating and maintenance procedures for the Dionex Easion. Operation Overview Figure 3-1 illustrates the basic steps for routine operation of the Dionex Easion. Turn on the Dionex Easion power Connect to Chromeleon...
• The conductivity detector begins reading the current conductivity. Connecting to Chromeleon If you are beginning operation of a Dionex Easion that has not been configured in a Chromeleon instrument, refer to the Chromeleon Help for setup instructions before proceeding.
3 • Operation and Maintenance Preparing the Regenerant The type of regenerant used with the Dionex Easion depends on the analysis. A dilute sulfuric acid regenerant is used for anion analyses; a tetrabutylammonium hydroxide (TBAOH) regenerant is used for cation analyses. Follow the instructions below to prepare either regenerant.
Page 38
6. Fill the bottle almost to the top with deionized water, and then place the bottle in the tray on the top of the Dionex Easion. NOTE To avoid staining the Dionex Easion, be careful not to spill TBAOH on the system.
3 • Operation and Maintenance ELUENT Bottle Out REGEN REGEN Bottle Out Bottle In Figure 3-2. Eluent and Regenerant Bottle Liquid Line Connections Priming the Pump Prime the pump if the eluent has been changed, the eluent line contains air, or the pump heads are dry (for example, after servicing).
Page 40
Prime option. 5. The pump will begin pumping at about 3 mL/min. Continue priming the Dionex Easion until no air bubbles are exiting the pump waste line. 6. Click the pump Off button on the Chromeleon ePanel.
Page 41
3 • Operation and Maintenance Priming the Eluent Line with a Syringe (Optional) A syringe can be used to facilitate priming when the eluent line is empty or the pump heads are completely dry. 1. Verify that the pump is turned off. 2.
Note that column manuals typically provide pressure specifications for systems that do not include a guard column. If a guard column is installed in the Dionex Easion, the system pressure will be 15% to 20% higher than the specification in the column manual.
3.6). Configuring Standby Mode The Dionex Easion can be configured to enter standby mode after a period of inactivity (defined as a period of time in which no data collection and no input from Chromeleon has occurred). In standby mode, the pump flow rate is reduced.
Dionex Easion Operator’s Manual • Analyze samples containing nitrite or sulfite as soon as possible. Nitrite oxidizes to nitrate, and sulfite to sulfate, thus increasing the measured concentrations of these ions in the sample. In general, samples that do not contain nitrite or sulfite can be refrigerated for at least one week with no significant changes in anion concentrations.
Page 45
3 • Operation and Maintenance diluting with eluent minimizes the effect of the water dip at the beginning of the chromatogram. If you dilute the sample with eluent, use eluent from the same lot to prepare the calibration standards, also. This is most important for fluoride and chloride, which elute near the water dip.
Dionex Easion Operator’s Manual 3.11 Processing Samples 3.11.1 Overview Samples can be run manually (one at a time) or grouped and run automatically (in batches). Figure 3-5 shows the typical steps for each type of sample processing. Manual Automated Sample Processing...
1. Complete the instructions in Section 3.2 through Section 3.10 prepare the Dionex Easion for operation and to prepare the sample for processing. 2. Load the sample into the injection valve sample loop with either the Dionex AS-DV (see Section 3.11.4) or a syringe (see...
Section 3.2 through Section 3.10 prepare the Dionex Easion for operation and to prepare the samples for processing. 2. If a Dionex AS-DV is installed: Prepare and fill sample vials and place them in the autosampler tray. Refer to the autosampler manual for detailed instructions.
Manually (with a Dionex AS-DV): Click Inject on the Chromeleon ePanel. Autosampler Setup Notes Follow these steps to verify that the Dionex Easion injection valve is controlled by the correct device. 1. Open the Chromeleon Instrument Configuration Manager. 2. In the Dionex Easion Properties dialog box, click the Inject Valve tab.
Section 3.11.4. Loading Samples with a Syringe (Push Method) 1. Verify that the sample loading port on the front of the Dionex Easion is connected to sample port on the injection valve (see S (5) Figure 3-6).
Page 51
Section 3.11.3. Loading Samples with a Vacuum Syringe (Pull Method) 1. Verify that the sample loading port on the front of the Dionex Easion is connected to sample port on the injection valve (see S (5) Figure 3-6).
Load on the Chromeleon ePanel (see Figure 2-10). 5. Insert a syringe into the sample loading port on the Dionex Easion front door and pull out the plunger to draw the sample into the injection valve. Draw at least five times the sample loop volume.
3 • Operation and Maintenance 3.12 Maintenance This section describes routine maintenance procedures for the Dionex Easion that users can perform. All other maintenance procedures must be performed by Thermo Fisher Scientific personnel. As Needed • Make fresh eluent. •...
Page 54
Dionex Easion Operator’s Manual Dionex Easion Preventive Maintenance Kit (P/N 088138) is available for this purpose. The kit provides parts and instructions for performing the procedure. • If a Dionex AS-DV is installed, perform the recommended annual preventive maintenance procedure. A Dionex AS-DV Preventive Maintenance Kit (P/N 072678) is available for this purpose.
4 • Troubleshooting This chapter is a guide to troubleshooting problems that may occur while operating the Dionex Easion. • For descriptions of error messages that may be displayed in the Chromeleon audit trail and suggestions for how to troubleshoot them, see Section 4.1.
Page 56
Abort page 51 Conductivity exceeds limit. If this error occurs when you turn on the Dionex Easion power, you can disregard it. If the error occurs at other times, it should be resolved. To troubleshoot: Refer to the suppressor manual for troubleshooting procedures.
Page 57
1. If a sequence is being executed, click Stop on the Chromeleon ePanel to cancel the current injection and stop the sequence. 2. Turn off the Dionex Easion power briefly and then restart. 3. Try to toggle the valve from Load to Inject by clicking Inject on the Chromeleon ePanel.
Page 58
For the Eluent Fill Volume to be accurate, you must enter the level each time the bottle is filled. The Dionex Easion does not automatically detect when the bottle is filled, nor when it is emptied. Doc. 155028-01 08/20...
Page 59
4 • Troubleshooting The system pressure is below the low pressure limit. If the system pressure falls below the minimum pressure limit for 0.5 second, Chromeleon stops the pump and this error message is displayed. The minimum pressure limit can be set to between 0 and 33.7 MPa (0 and 4900 psi) in the Chromeleon Instrument Configuration Manager.
Check that the lines are not crimped or otherwise blocked. Also, if the blocked line is a waste line, make sure it is not elevated at any point after it exits the Dionex Easion. If a line is blocked, replace it (see Section 5.2).
4 • Troubleshooting • Leaking Dionex CRS 500 Refer to the suppressor manual for troubleshooting procedures. • Leaking injection valve 1. Make sure the liquid line connections to the valve are tight. Replace any damaged fittings. Refer to Installation of Dionex Liquid Line Fittings (Document No.
Dionex Easion Operator’s Manual • No communication between Dionex Easion and Chromeleon ( Link on rear panel fails to light) The USB cable is not connected correctly. For connection instructions, refer to the Dionex Easion installation instructions. No Flow •...
4 • Troubleshooting • Flow rate through columns too high Check the pump flow rate. • Clogged column bed supports Refer to the column manual for troubleshooting procedures. • Contaminated columns Clean the columns as instructed in the column manual. •...
Dionex Easion Operator’s Manual Nonreproducible Peak Height or Retention Time • Column overloading Dilute the sample (see Section 3.10.3). • Liquid leaks Locate and eliminate the leaks (see Section 4.2). • Incomplete or imprecise filling of the sample loop 1. Fill the sample loop until excess sample exits the waste line.
Chromeleon Help.) If the conductivity reading with the dummy cell is outside the tolerance range, the electronics are malfunctioning. Contact Technical Support for Dionex products for assistance. The Dionex Easion electronics components cannot be serviced by the user. 4.12 High Detector Output •...
Page 66
If the ambient temperature does not meet the specification of 10 to 35 °C (50 to 95 °F), verify that air conditioning and heating vents are directed away from the Dionex Easion and that the Dionex Easion front door is closed. •...
Page 67
• DCR waste backpressure tubing not installed (or installed incorrectly) Verify that the DCR waste backpressure tubing is installed on the end of the suppressor waste line. For more information, refer to the Dionex Easion installation instructions. Doc. 155028-01 08/20...
5 • Service This chapter describes Dionex Easion service and repair procedures that users can perform. Procedures not included here, including electronics-related repair procedures, must be performed by Thermo Fisher Scientific personnel. For assistance, contact Technical Support for Dionex products before proceeding: •...
Click the pump On or Off button to turn the pump on or off, respectively. • Click the Log Serial button to record the Dionex Easion serial number in the Chromeleon audit trail. • The Calibration or Diagnostic box is green while a calibration or diagnostic procedure is running.
Audit Trail The Chromeleon audit trail displays an account of every event (including errors, status messages, and operational events) that occurs during Dionex Easion operation. 5.1.2 Calibrating the Pressure Transducer 1. To remove any air in the injection valve loop, toggle the injection valve position a few times by clicking Load and Inject on the Chromeleon ePanel.(You should hear a click when the valve position...
Page 72
Pressure Transducer Waste Valve Figure 5-1. Dionex Easion Interior Components 6. Close the waste valve. 7. Under Pressure Transducer, click Calibrate Slope. A panel with instructions and command buttons appears. 8. Follow the instructions on the panel to complete the transducer slope calibration procedure.
5 • Service 5.1.3 Calibrating the Flow Rate Calibrate the flow rate if the Operational Qualification or Performance Qualification flow rate accuracy and precision test fails. Items needed: • 0.076-mm (0.003-in) ID yellow PEEK tubing (P/N 049715) to create 14 ± 2 MPa (2000 ± 300 psi) of backpressure (if needed) •...
Dionex Easion Operator’s Manual Replacing Tubing and Fittings The Dionex Easion is plumbed with the tubing and tubing assemblies listed below. Tubing Size and Type Used for... 0.125-mm (0.005-in) ID PEEK, red Connection from pump pulse damper to (P/N 044221) pressure transducer 0.25-mm (0.010-in) ID PEEK, black...
Isolating a Restriction in the Liquid Plumbing A restriction in the liquid plumbing will cause excessive system backpressure. 1. Begin pumping eluent through the system (including the columns). 2. Follow the Dionex Easion flow schematic (see Figure 5-2) to work backward through the system, beginning at the cell exit.
1. Click the pump Off button on the Chromeleon ePanel. 2. Open the Dionex Easion Sample Loop front door. 3. Disconnect the sample loop from ports on the injection valve (see Figure 5-3). Figure 5-3. Dionex Easion Injection Valve Doc. 155028-01 08/20...
(see Section 5.7). To remove the old check valves: 1. Turn off the Dionex Easion power switch. 2. Disconnect the power cord. 3. Put on standard disposable laboratory rubber gloves (powder-free, particle- free, and oil-free) before disassembling the pump head.
Page 79
5 • Service Outlet Check Valve Secondary Primary Pump Head Pump Head Inlet Check Valve Figure 5-4. Pump Heads To install the new check valves: 1. Inspect the new inlet check valve Inlet Check Valve Outlet Check Valve assembly (P/N 045722) to verify that the 1/4-28 Port 10-32 Port double-hole end of the cartridge is...
6. Prime the pump (see Section 3.6). 7. When the Dionex Easion is at operating pressure, check for leaks from the check valves. Tighten a check valve a little more only if it leaks. Cleaning Pump Check Valves If a check valve leaks or is dirty, both check valves should be replaced (see Section 5.6).
2. After rinsing, close the waste valve. To remove the pump head and piston: 1. Turn off the Dionex Easion power switch. 2. Disconnect the power cord. 3. Put on standard disposable laboratory rubber gloves (powder-free, particle- free, and oil-free) before disassembling the pump head.
Page 82
Dionex Easion Operator’s Manual Vermeiden Sie Seitwärtsbewegungen, wenn Sie den Pumpenkopg vom Kolben lösen. Andernfalls kann der Kolben brechen. 7. Place the pump head (front end down) on a clean work surface. 8. The piston does not come off as part of the pump head assembly because it is captured by a magnetic retention system.
Page 83
5 • Service Figure 5-7. Secondary Pump Head To install the new backup piston seal: 1. Lift the spacer off the pump head. 2. Remove the seal holder from the spacer, using one of the following methods: • With the seal holder facing up, cup the spacer Notch in your hand and blow clean laboratory air at the spacer to dislodge the seal holder.
Page 84
Dionex Easion Operator’s Manual 4. To install the new backup piston seal: Grooved side of backup a. Hold the new backup piston seal piston seal (P/N 063382) with the grooved side facing down, and insert the shaft of the piston through the seal.
Page 85
5 • Service 3. If this is the primary pump head, install a 10-32 fitting plug (P/N 042772) on the outlet check valve. Tighten the plug. If this is the secondary pump head, install a 10-32 fitting plug (P/N 042772) in both the inlet and outlet ports.
Page 86
Dionex Easion Operator’s Manual 5. To seat the seal, push down on the spacer until it is flush with the pump head. A clicking sound indicates that the seal is correctly seated. 6. Remove the spacer and piston and verify that the seal is flush with the surface of the pump head.
Tighten the screws evenly. To complete the procedure: 1. Reconnect all liquid lines to the pump head. 2. Close the waste valve knob. 3. Turn on the Dionex Easion power switch. 4. Prime the pump (see Section 3.6). Replacing a Pump Piston Continued leaking of the piston seal after installation of a new seal (assuming the pump head is tight) indicates a dirty, scratched, or broken piston.
Dionex Easion Operator’s Manual 5.10 Replacing the Waste Valve Seal A damaged seal causes leakage around the base of the waste valve knob. 1. Turn off the Dionex Easion power switch. 2. To remove the waste valve from the pump head (see Figure...
1. Click the pump Off button on the Chromeleon ePanel. 2. Open the front door of the Dionex Easion. 3. Disconnect each liquid line connected to the injection valve. 4. Follow the instructions provided in the Injection Valve Rebuild Kit to replace the rotor seal, isolation seal, and stator face.
Page 90
CELL OUT 11. Attach the line to the port on the suppressor. ELUENT OUT ELUENT OUT 12. Close the Dionex Easion front door. 13. Reconnect the power cord and turn on the Dionex Easion power. Doc. 155028-01 08/20...
Refer to the suppressor manual for guidance about when to replace a suppressor. 1. Click the pump Off button on the Chromeleon ePanel. 2. Open the front door of the Dionex Easion. 3. Disconnect the four eluent and regenerant lines from the suppressor.
Page 92
Dionex Easion Operator’s Manual 3. The fuse drawer is located above the power switch (see Figure 5-11). A small Fuse Drawer tab locks the fuse drawer in place. Using a small screwdriver, press the tab in and then up to release the fuse drawer.
A • Specifications A.1 Electrical Main Power 100 to 240 Vac, 50 to 60 Hz, autoranging Maximum input power: 90 W Maximum line draw: 1.5 A Fuses Two fast-blow IEC 127 fuses rated 3.15 A (P/N 954745) A.2 Physical Dimensions Height without bottles: 33 cm (13 in) Height with bottles and cap tubing: 60 cm (24 in) Width: 24 cm (9.5 in)
Front Panel LEDs Power Indicates when the power is on. Ready Indicates when the Dionex Easion is ready to acquire data; flashes if the system check fails. Indicates when the Dionex Easion is running/acquiring data; flashes if an error occurs.
Factory-calibrated before shipment to the user; calibrated in the field with memory retention and transfer Control and Provided by Chromeleon software; communication with the Data Evaluation Dionex Easion is via USB (Universal Serial Bus) Cell Body PEEK Cell Electrodes Passivated 316 stainless steel 1 L...
A • Specifications A.9 Autosampler Automation Dionex AS-DV Autosampler (or third-party autosampler) Using Autosampler Sequential/ Dependent on autosampler capabilities Simultaneous Injection Inline Sample Optional with Dionex CRD 200 or Dionex CRD 300 Degassing Inline Filtration Dionex AS-DV Autosampler or inline filter High Automation Conditionals using Chromeleon and post-run features Flexibility...
Page 98
Dionex Easion Operator’s Manual System Status Standard feature Virtual Channels Power Failure Standard feature Protection Sequential Standard feature for independent dual-system analysis Injection System Trigger Standard feature Commands and Conditionals Daily Audit Trail Standard feature Sample Audit Standard feature Trail...
B • Reordering Information Part Number Item Pump 067701 Primary pump head assembly 067703 Secondary pump head assembly 045722 Inlet check valve assembly, 1/4-28 045721 Outlet check valve assembly, 10-32 055870 Piston seal 014895 O-ring 063382 Backup piston seal 067706 Seal holder 052840 Piston...
C • TTL and Relay Control A 12-pin connector strip for TTL/relay control is located on the Dionex Easion rear panel. The connector provides two relay outputs, two TTL outputs, and four TTL inputs (see Figure C-1). Pin Function Description...
When attaching wires to the connector plug, be careful not to allow stray strands of wire to short to the adjoining position on the connector. b. Plug the connector into the 12-pin connector on the Dionex Easion rear panel. Doc. 155028-01 08/20...
Normal Edge Thermo Scientific Dionex modules. If the device connected to the Dionex Easion does not send a normal edge signal, select the appropriate control mode. Refer to the documentation provided with the controlling device, as well as the information below, to select the correct type.
Page 104
Dionex Easion Operator’s Manual • Normal Edge: In normal edge Action Off or No Effect operation, the negative (falling) edge of +5 V a signal turns on the function. For NORMAL EDGE example, for the Load/Inject function, +0 V the negative edge switches the injection Action On valve position to Load.
1. Open the Chromeleon Instrument Configuration Manager. 2. In the Dionex Easion Properties dialog box, click the State Devices tab. 3. Select the relay output name (Pump_ECD_Relay_1 or Pump_ECD_ Relay_2) and press the F2 key.
Page 106
Dionex Easion Operator’s Manual 2. Expand the list of commands under Pump_ECD and select the TTL or relay output name. 3. Execute the command. To program control of TTL and relay outputs: Include commands for control of the TTL and relay outputs in the Chromeleon instrument method.
D • FAQ D.1 How do I connect to the Dionex AS-DV? For instructions on how to connect the Dionex Easion to the Dionex AS-DV, refer to the autosampler operator’s manual. D.2 Why are the retention times moving? Retention times can shift if the pump flow is erratic, or if the column or eluent is contaminated.
Click the pump Off button on the Chromeleon ePanel (see Figure 2-10). –or– Turn off the power switch on the Dionex Easion rear panel (see Figure 2-3). D.7 How do I store columns? Columns should be stored in eluent. Refer to the column manual for complete instructions.
D • FAQ D.9 How do I clean a column? Refer to the troubleshooting section of the column manual. D.10 Why is the conductivity high? Possible reasons for high conductivity include: • The suppressor needs regeneration. Refer to the troubleshooting section of the suppressor manual.
E • Introduction to Ion Chromatography (IC) The Dionex Easion performs isocratic ion analyses using suppressed conductivity detection. An ion chromatography system such as the Dionex Easion typically consists of a liquid eluent, a high-pressure pump, a sample injector, a separator column, a chemical suppressor, and a conductivity cell.
Page 112
• As the eluent and sample are pumped through the separator column, the sample ions are separated. In the Dionex Easion, the mode of separation is called ion exchange. This mode is based on the premise that different sample ions migrate through the IC column at different rates, depending upon their interactions with the ion exchange sites.
F • Glossary Analytical Column Synonymous with Separator Column. Band Spreading The broadening of the sample band as it travels through the column. Band spreading can also occur in the injection valve, detector cell, and interconnecting tubing. Calibration Curve A graph showing detector response in peak height or area versus analyte concentration.
Page 114
Dionex Easion Operator’s Manual Channeling The preferential flow of liquid along more open, less resistant paths through the column packing. This causes Band Spreading. Column Efficiency (N) A measure of the narrowness of analyte bands as they elute from the column.
Page 115
F • Glossary % Crosslink Divinylbenzene content in a polystyrene/divinylbenzene (PS-DVB) resin; this contributes to the mechanical strength of the resin and determines chromatographic properties. Equivalent Conductivity ( The contribution of an ionic species to the total conductivity of a solution as measured in a standard cell having electrodes 1 cm in area and exactly 1 cm apart.
Page 116
Dionex Easion Operator’s Manual Regenerant A dilute acid or base which converts ion exchange sites in the suppressor back to the form which will suppress the eluent conductivity. Resin See Ion-Exchange Resin. Resolution (R) A measure of the separation between two sample components. It is expressed as the ratio of the distance between the two peak maxima to the mean value of the peak width at the baseline.
Page 117
F • Glossary Void Volume (V The volume occupied by the eluent in a packed column. This volume includes the volume between the injection valve and the column, as well as between the column and the detector cell. Unretained components are eluted in the void volume.
Need help?
Do you have a question about the Dionex Easion and is the answer not in the manual?
Questions and answers