Thermo Scientific Dionex ICS Series Operator's Manual page 10

Vwd variable wavelength detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

Dionex VWD Operator's Manual
Two light sources, a deuterium lamp for ultraviolet detection and a tungsten
lamp for visible and near-infrared wavelength detection, provide a
wavelength detection range from 190 to 900 nm. For details about the lamps
and wavelength detection ranges, see
The four-channel detector measures at up to four wavelengths, by indexing
between wavelengths several times per second.
To suppress higher-order radiation, two optical filters are inserted
(automatically) into the light path (see
The wavelength accuracy can be verified via the built-in holmium oxide filter
(see
Section
The detector is controlled with a PC (personal PC) running Microsoft
Windows Vista
Chromeleon
Chromatography Data System. Chromeleon 7 and Chromeleon 6.8 also
provide data acquisition and data processing functions.
NOTE Thermo Scientific Dionex Chromeleon
Various monitoring, diagnostic, and calibration features are provided for
system wellness and predictive performance (see
2
5.2).
®
®
or Windows
XP and either the Thermo Scientific Dionex
®
7 Chromatography Data System or Chromeleon
can provide real-time control and monitoring of the
detector, but does not include data management
capabilities.
Section
2.2.4.
Section
2.3).
®
Xpress software
Section
2.6.2).
®
®
6.8
Doc. 065377-01 1/12