GeoMx DSP Instrument User Manual
Administration Menu
In the event that you received an error when setting up the archive, check the following
specifications and tips:
Ensure that your institute network is set up for SMB3 file sharing.
l
Enter the Remote Server path as \\\\123.45.6.78\\FolderName using the server IP address
l
and your designated folder.
Try the username in different formats, such as username, username@domain, or
l
username@domain.edu
Passwords may not have the characters $ ! /
l
Confirm that the credentials to access the folder have reading and writing permission.
l
Nested folders may cause problems – try to set the Archive folder in the root of the server as
l
shown in above example.
A network load balancer may be an issue – contact GeoMx Support to discuss
l
workarounds.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
80
MAN-10152-01 for software v2.5