Unload Plate And Slides - nanoString GeoMx DSP User Manual

Instrument
Table of Contents

Advertisement

MAN-10152-01 for software v2.5

Unload Plate and Slides

Unload Plate and Slides
Once ROI collection is complete, the system will initiate an Instrument Cleaning process and
end with the workflow marker at the top of the DSP Control Center at Complete
47). Each scan image will have the status Collection Complete displayed.
This Collection Complete phase is an idle software state and is a good time to Change
Reagents.
Choose the New Data Collection button to begin a new collection. You will have the opportunity
to exchange the slides and/or the collection plate. Choose the Remove Slides and Microplate
button to remove these materials from the instrument, without initiating a new run.
Removing the collection plate
Follow the prompts in the GeoMx DSP Control Center to
remove the collection plate from the instrument
Figure
48).
After the collection plate has been removed from the DSP
instrument, it should be immediately prepared for readout
by NGS (see
GeoMx DSP NGS Readout User Manual
(MAN-10153)) or nCounter (see
User Manual
(MAN-10089)).
If storing plate before processing, seal plate with
adhesive foil to prevent contamination. Store plate following these guidelines:
If stored 24 hours or less: store at 4°C
l
If stored between 24 hours and 30 days: store at -20°C
l
If stored longer than 30 days: store at -80°C
l
IMPORTANT:
Deviating from the safe storage guidelines may result in reductions in data
quality.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
Figure 47: Workflow marker at Collection Complete
GeoMx nCounter Readout
GeoMx DSP Instrument User Manual
(see
Figure 48: Unloading the collection
plate
(see Figure
59